Comparative Study
. 2013 Oct;21(10):2200-5.
doi: 10.1007/s00167-013-2625-6. Epub 2013 Aug 13.

Radiographic accuracy in TKA with a CT-based patient-specific cutting block technique

Affiliations
Free article
Comparative Study

Radiographic accuracy in TKA with a CT-based patient-specific cutting block technique

P P Koch et al. Knee Surg Sports Traumatol Arthrosc. 2013 Oct.
Free article

Abstract

Purpose: Patient-specific instrumentation (PSI) technology for the implantation of total knee arthroplasty (TKA) has a rising interest in the orthopaedic community. Data of PSI are controversially discussed. The hypothesis of this paper is that the radiological accuracy of CT-based PSI is similar to the one of navigated TKA published in the literature.

Methods: Since 2010, all 301 consecutively performed PSI TKAs (GMK MyKnee©) were included in this study. The radiological assessment consisted in a preoperative and postoperative standard X-ray and long-standing X-ray. Changes from the planned to the definitively implanted component size were documented. Postoperative analysis included limb alignment and position of femoral and tibial components (for varus/valgus and flexion or tibial slope).

Results: The postoperative average hip-knee-ankle angle was 180.1° ± 2.0°. In the frontal plane a total of 12.4 % of outliers >3°, for the tibial components 4.1 % of outliers >3° and for the femoral components 4.8 % of outliers >3° were measured. A total of 12.3 % of outliers for posterior tibial slope and 9 % of outliers >3° for the femoral flexion were noted. 10.8 % of the 602 planned size components were adapted intraoperatively.

Conclusion: Although it is still unknown which limb axis is the correct one for the best clinical result, a technology providing the aimed axis in a most precise way should be chosen. Comparing the outcome of the current study with the data from the literature, there does not seem to be any difference compared to computer-assisted surgery.

Level of evidence: IV.