Randomized Controlled Trial
. 2013 Oct;21(10):2213-9.
doi: 10.1007/s00167-013-2623-8. Epub 2013 Aug 14.

Rotation in total knee arthroplasty: no difference between patient-specific and conventional instrumentation

Affiliations
Randomized Controlled Trial

Rotation in total knee arthroplasty: no difference between patient-specific and conventional instrumentation

Sébastien Parratte et al. Knee Surg Sports Traumatol Arthrosc. 2013 Oct.

Abstract

Purpose: It was our hypothesis that patient-specific instrumentation (PSI) can improve the accuracy of the rotational alignment in TKA based on the concept of the system and on the potential to clearly identify pre-operatively during planning the classical anatomical landmarks that serve as references to set-up the rotation both for the femur and tibia.

Materials and methods: In this prospective comparative randomized study, 40 patients (20 in each group) operated in our institution between September 2012 and January 2013 by the 2 senior authors were included. Randomization of patients into one of the two groups was done by the Hospital Informatics Department with the use of a systematic sampling method. All patients received the same cemented high-flex mobile bearing TKA. In the PSI group, implant position was compared to the planed position using previously validated dedicated software. The position of the implants (frontal and sagittal) was compared in the 2 groups on standard X-rays, and the rotational position was analysed on post-operative CT-scan.

Results: 90 % of the patients add <2° or mm of difference between the planned position of the implants and the obtained position, except for the tibial rotation where the variations were much higher. Mean HKA was 179° (171-185) in the PSI group with 4 outliers (2 varus: 171° and 172°:184° and 185°) and 178.3° with 2 outliers (171° and 176°) in the control group. No difference was observed between the two groups concerning the frontal and sagittal position of the implants on the ML and AP X-rays. No significant difference of femoral rotation was observed between the two groups with a mean of 0.4° in the PSI group and 0.2° in the control group (p: n.s). Mean tibial rotation was 8° of internal rotation in the PSI group and 15° of internal rotation in the standard group (p: n.s).

Conclusion: Based on our results, we were unable to confirm our hypothesis as PSI cannot improve rotation in TKA. More work needs to be done to more clearly define the place of PSI in TKA, to keep on improving the accuracy of the system and to better define the individual targets in TKA in terms of frontal, sagittal and rotational positioning of the implant for each patient.

Level of evidence: Prospective comparative randomized study, Level II.

PubMed Disclaimer

References

    1. Knee Surg Sports Traumatol Arthrosc. 1997;5(4):251-7 - PubMed
    1. Knee Surg Sports Traumatol Arthrosc. 2013 Oct;21(10):2370-4 - PubMed
    1. J Arthroplasty. 1997 Sep;12(6):677-82 - PubMed
    1. J Bone Joint Surg Am. 2010 Jul 7;92(7):1635-9 - PubMed
    1. Clin Orthop Relat Res. 2012 Mar;470(3):895-902 - PubMed

Publication types

MeSH terms

LinkOut - more resources