Clinical Trial
. 2013 Oct;21(10):2271-80.
doi: 10.1007/s00167-013-2621-x. Epub 2013 Aug 15.

Accurate alignment and high function after kinematically aligned TKA performed with generic instruments

Affiliations
Clinical Trial

Accurate alignment and high function after kinematically aligned TKA performed with generic instruments

Stephen M Howell et al. Knee Surg Sports Traumatol Arthrosc. 2013 Oct.

Abstract

Purpose: Performing kinematically aligned total knee arthroplasty (TKA) with generic instruments is less costly than patient-specific instrumentation; however, the alignment and function with this new technique are unknown.

Methods: One hundred and one consecutive patients (101 knees) treated with kinematically aligned TKA, implanted with use of generic instruments, were prospectively followed. The medial collateral ligament was not released. The lateral collateral ligament was released in the 17 % of patients with a fixed valgus deformity. Six measures of alignment were categorized from a scanogram of the extremity, an axial scan of the knee, and an intraoperative measurement. Both the Oxford Knee and WOMAC™ scores were assessed as function. High function was a mean Oxford Knee score >41.

Results: The frequency that patients were categorized as in-range was 93 % for the mechanical alignment of the limb (0° ± 3°), 94 % for the joint line (-3° ± 3°), 57 % for the anatomic axis of the knee (-2.5° ± -7.4° valgus), 4 % for the varus-valgus rotation of the tibial component (≤0° valgus), 98 % for the rotation of the tibial component with respect to the femoral component (0° ± 10°), and 94 % for the intraoperative change in the anterior-posterior distance of the tibia with respect to the femur at 90° of flexion (0 ± 2 mm). The mean OKS score was 42, and WOMAC™ score was 89. For each alignment, the function was the same for patients categorized as an outlier or in-range.

Conclusions: The authors prefer the use of generic instruments to perform kinematically aligned TKA in place of mechanically aligned TKA because five of six alignments were accurate and because high function was restored regardless of whether patients had an alignment categorized as an outlier or in-range.

Level of evidence: IV.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Clin Orthop Relat Res. 2013 Mar;471(3):1000-7 - PubMed
    1. Knee Surg Sports Traumatol Arthrosc. 2010 Mar;18(3):388-93 - PubMed
    1. J Bone Joint Surg Am. 2011 Sep 7;93(17):1588-96 - PubMed
    1. Clin Orthop Relat Res. 2011 Aug;469(8):2346-55 - PubMed
    1. BMC Musculoskelet Disord. 2010 Mar 25;11:57 - PubMed

Publication types

MeSH terms