Comparison of matrix-assisted laser desorption ionization-time-of-flight mass spectrometry assay with conventional methods for detection of IMP-6, VIM-2, NDM-1, SIM-1, KPC-1, OXA-23, and OXA-51 carbapenemase-producing Acinetobacter spp., Pseudomonas aeruginosa, and Klebsiella pneumoniae
- PMID: 23993215
- DOI: 10.1016/j.diagmicrobio.2013.07.005
Comparison of matrix-assisted laser desorption ionization-time-of-flight mass spectrometry assay with conventional methods for detection of IMP-6, VIM-2, NDM-1, SIM-1, KPC-1, OXA-23, and OXA-51 carbapenemase-producing Acinetobacter spp., Pseudomonas aeruginosa, and Klebsiella pneumoniae
Erratum in
- Diagn Microbiol Infect Dis. 2014 Oct;80(2):170
Abstract
Matrix-assisted laser desorption ionization-time-of-flight mass spectrometry assay was able to detect carbapenemase producers, including SIM-1 or OXA-51, within 4 hours using 20 μL of 0.5 g/L ertapenem solution as a substrate. This assay is more rapid and accurate than the modified Hodge test and 3-dimensional extract bioassay. Hence, it can be used an alternative test to identify carbapenemase-mediated carbapenem resistance in Gram-negative bacteria.
Keywords: Acinetobacter spp; Carbapenemase; MALDI-TOF MS; OXA; SIM-1.
© 2013.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous