Biomechanical comparison of syndesmotic injury fixation methods using a cadaveric model
- PMID: 24019217
- DOI: 10.1177/1071100713503816
Biomechanical comparison of syndesmotic injury fixation methods using a cadaveric model
Abstract
Background: There is growing interest in suture-button devices for syndesmosis injury, which are intended to offer less rigid fixation than screw fixation.
Methods: The fixation strength with 2 different suture-button devices, ZipTight and TightRope, were compared using 5 cadaveric leg pairs (n = 10). In an additional 5 pairs (n = 10), ZipTight was compared to 3.5 mm quadricortical screw fixation. Ankle motion was measured intact, then following simulated syndesmosis injury and fixation. Cyclic loads (peak 750 N, 7.5 Nm) were applied. Finally, external rotation to failure was measured and failure mode was documented.
Results: Range of motion increased after simulated injury and fixation with all devices (max 14.5 degrees). In all groups, diastasis remained below 1.0 mm intact and below 2.0 mm during cyclic loading. Compared to intact, under load to failure, diastasis with ZipTight devices increased by 4.7 ± 1.3 mm and 7.6 ± 4.3 mm, with TightRope, 6.3 mm, and screw construct, 1.3 mm. ZipTight specimens rotated approximately 80 ± 22 degrees before failure, TightRope, 67 ± 13 degrees, screw constructs, 76 ± 27 degrees. Mean failure torque was between 22.2 ± 6.9 Nm and 28.1 ± 12.7 Nm for ZipTight, compared to 32.9 ± 8.0 Nm for TightRope (P = .07), and 30.1 ± 9.6 Nm for screw constructs (P = .03). The majority of suture-button constructs failed by fibular fracture (ZipTight = 6, TightRope = 4), the remaining by device pull-through (ZipTight = 3, TightRope = 1) and loosening (ZipTight = 1). Conversely, 3 of screw-fixed specimens failed by device failure, 2 from bone fracture.
Conclusion: Suture-button devices provided torsional strength below that of screw fixation. However, all devices may provide failure torques well above 20 Nm, exceeding likely torques applied in casts during healing.(1,2,4) CLINICAL RELEVANCE: Suture-button devices appear to have provided adequate fixation strength for syndesmosis injuries.
Keywords: ankle fixation; screw fixation; suture-button; syndesmosis injury.
Similar articles
-
Suture-button versus screw fixation in a syndesmosis rupture model: a biomechanical comparison.Foot Ankle Int. 2009 Apr;30(4):346-52. doi: 10.3113/FAI.2009.0346. Foot Ankle Int. 2009. PMID: 19356360
-
Biomechanical Comparison of 3 Current Ankle Syndesmosis Repair Techniques.Foot Ankle Int. 2017 Feb;38(2):200-207. doi: 10.1177/1071100716666278. Epub 2016 Oct 1. Foot Ankle Int. 2017. PMID: 27672016
-
Suture-button versus screw fixation of the syndesmosis: a biomechanical analysis.Foot Ankle Int. 2010 Jan;31(1):69-75. doi: 10.3113/FAI.2010.0069. Foot Ankle Int. 2010. PMID: 20067726
-
Biomechanical comparison of suture-button, bioabsorbable screw, and metal screw for ankle syndesmotic repair: A meta-analysis.Foot Ankle Surg. 2021 Feb;27(2):117-122. doi: 10.1016/j.fas.2020.03.008. Epub 2020 Apr 27. Foot Ankle Surg. 2021. PMID: 32466876 Review.
-
Comparison of suture button fixation and syndesmotic screw fixation in the treatment of distal tibiofibular syndesmosis injury: A systematic review and meta-analysis.Int J Surg. 2018 Dec;60:120-131. doi: 10.1016/j.ijsu.2018.11.007. Epub 2018 Nov 12. Int J Surg. 2018. PMID: 30439535
Cited by
-
A biomechanical comparison study of a modern fibular nail and distal fibular locking plate in AO/OTA 44C2 ankle fractures.J Orthop Surg Res. 2016 Sep 15;11(1):100. doi: 10.1186/s13018-016-0435-5. J Orthop Surg Res. 2016. PMID: 27628500 Free PMC article.
-
Biomechanics of the Distal Tibiofibular Syndesmosis: A Systematic Review of Cadaveric Studies.Foot Ankle Orthop. 2021 Jun 4;6(2):24730114211012701. doi: 10.1177/24730114211012701. eCollection 2021 Apr. Foot Ankle Orthop. 2021. PMID: 35097448 Free PMC article. Review.
-
Effect of a Controlled Ankle Motion Walking Boot on Syndesmotic Instability During Weightbearing: A Cadaveric Study.Orthop J Sports Med. 2019 Aug 20;7(8):2325967119864018. doi: 10.1177/2325967119864018. eCollection 2019 Aug. Orthop J Sports Med. 2019. PMID: 31457067 Free PMC article.
-
Effects and Anti-rotation Stabilization of the Non-bridging External Fixation for Pronation-Abduction Stage III Ankle Fracture: A Cadaveric Study.Biomed Res Int. 2021 May 8;2021:9966344. doi: 10.1155/2021/9966344. eCollection 2021. Biomed Res Int. 2021. PMID: 34046502 Free PMC article.
-
Proximal Tibiofibular Dislocation Repaired with Syndesmotic Suture Buttons: A Case Report.J Orthop Case Rep. 2018 Sep-Oct;8(5):82-85. doi: 10.13107/jocr.2250-0685.1224. J Orthop Case Rep. 2018. PMID: 30740384 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical