Echocardiographic predictors of cardiac erosion after Amplatzer septal occluder placement
- PMID: 24038861
- DOI: 10.1002/ccd.25175
Echocardiographic predictors of cardiac erosion after Amplatzer septal occluder placement
Abstract
The risk of erosion after Amplatzer septal occluder (ASO) device placement in atrial septal defects is well described. Aortic rim deficiency and use of over-sized device increase the risk of erosion. This study attempts to describe device characteristics, anatomical features and echocardiographic predictors that increase the risk of erosion.
Methods: From 2005 through 2012, 12 new cases, with nine confirmed and three suspected device erosions where pre-procedural, intra-procedural, and/or post-procedural echocardiograms were available and, were reviewed. Following parameters were evaluated: ASD location (high or low), rims deficiency and consistency, septal mal-alignment, dynamic nature of the defect; device edge relationship toward the transverse sinus (TS), atrial free wall tenting and the size of the defect compared with the size of the device used for closure.
Results: We found poor posterior rim consistency, aortic rim absence (in multiple views) and absent aortic rim at O degree in 100% of the patients. Septal mal-alignment and dynamic ASD was present in nearly 50% of the cases. The device was over-sized in three patients only. A 26-mm device was the most common device that resulted in erosion. In cases, where patient had experienced bloody pericardial effusion and the device was in place, device tenting in the TS was observed. Surgical explantation of the device confirmed presence of erosion in all cases.
Conclusion: Aortic rim absence in multiple views, poor posterior rim consistency, septal mal-alignment, and dynamic ASD appear to be factors where erosion risk increases significantly. A thorough assessment of the device edge by echocardiography in short-axis may show device tenting of the atrial free wall into the TS. This finding should be a strong indictor to recommend surgical removal of device after occurrence of pericardial effusion.
Keywords: complications pediatric cath/intervention; patent foramen ovale/atrial septal defect; pediatric interventions.
Copyright © 2013 Wiley Periodicals, Inc.
Comment in
-
Look before (and after) you plug: moving slowly toward safer selection and management of patients at risk of device erosion.Catheter Cardiovasc Interv. 2014 Jan 1;83(1):93-4. doi: 10.1002/ccd.25288. Catheter Cardiovasc Interv. 2014. PMID: 24347119 No abstract available.
Similar articles
-
Growth of the atrial septum after Amplatzer device closure of atrial septal defects in young children.Catheter Cardiovasc Interv. 2015 Nov 15;86(6):1041-7. doi: 10.1002/ccd.26041. Epub 2015 May 27. Catheter Cardiovasc Interv. 2015. PMID: 26013563
-
Very late erosion of Amplatzer septal occluder device presenting as pericardial pain and effusion 8 years after placement.Catheter Cardiovasc Interv. 2013 Oct 1;82(4):E592-4. doi: 10.1002/ccd.24755. Epub 2013 Mar 18. Catheter Cardiovasc Interv. 2013. PMID: 23172721
-
Relative Risk Factors for Cardiac Erosion Following Transcatheter Closure of Atrial Septal Defects: A Case-Control Study.Circulation. 2016 May 3;133(18):1738-46. doi: 10.1161/CIRCULATIONAHA.115.019987. Epub 2016 Mar 21. Circulation. 2016. PMID: 27002094
-
Percutaneous atrial septal occluder devices and cardiac erosion: a review of the literature.Catheter Cardiovasc Interv. 2012 Aug 1;80(2):157-67. doi: 10.1002/ccd.24347. Epub 2012 May 2. Catheter Cardiovasc Interv. 2012. PMID: 22552868 Review.
-
Infective endocarditis after device closure of atrial septal defects: Case report and review of the literature.Catheter Cardiovasc Interv. 2017 Feb 1;89(2):324-334. doi: 10.1002/ccd.26784. Epub 2016 Sep 19. Catheter Cardiovasc Interv. 2017. PMID: 27641431 Review.
Cited by
-
Very late erosion of Amplatzer occluder device resulting in Cardiac tamponade after 15 years.Clin Res Cardiol. 2018 Jun;107(6):527-529. doi: 10.1007/s00392-018-1222-6. Epub 2018 Feb 26. Clin Res Cardiol. 2018. PMID: 29484456 No abstract available.
-
Delayed Left Atrial Perforation Associated with Erosion After Device Closure of an Atrial Septal Defect.Korean J Thorac Cardiovasc Surg. 2017 Apr;50(2):110-113. doi: 10.5090/kjtcs.2017.50.2.110. Epub 2017 Apr 5. Korean J Thorac Cardiovasc Surg. 2017. PMID: 28382270 Free PMC article.
-
Trends in surgical and catheter interventions for isolated congenital shunt lesions in the UK and Ireland.Heart. 2019 Jul;105(14):1103-1108. doi: 10.1136/heartjnl-2018-314428. Epub 2019 Feb 16. Heart. 2019. PMID: 30772822 Free PMC article.
-
Transcatheter closure of atrial septal defect: principles and available devices.J Thorac Dis. 2018 Sep;10(Suppl 24):S2909-S2922. doi: 10.21037/jtd.2018.02.19. J Thorac Dis. 2018. PMID: 30305951 Free PMC article. Review.
-
A Novel Patent Foramen Ovale Occluder With Left Clamp Arms: Preclinical Experiment and First-in-Human Study.JACC Basic Transl Sci. 2025 Apr;10(4):422-433. doi: 10.1016/j.jacbts.2024.11.012. Epub 2025 Feb 5. JACC Basic Transl Sci. 2025. PMID: 40306851 Free PMC article.
MeSH terms
Supplementary concepts
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials