. 2013 Dec;82(12):2253-7.
doi: 10.1016/j.ejrad.2013.08.035. Epub 2013 Aug 30.

Radiation exposure in CT-guided interventions

Affiliations

Radiation exposure in CT-guided interventions

Roman Kloeckner et al. Eur J Radiol. 2013 Dec.

Abstract

Purpose: To investigate radiation exposure in computed tomography (CT)-guided interventions, to establish reference levels for exposure, and to discuss strategies for dose reduction.

Materials and methods: We analyzed 1576 consecutive CT-guided procedures in 1284 patients performed over 4.5 years, including drainage placements; biopsies of different organs; radiofrequency and microwave ablations (RFA/MWA) of liver, bone, and lung tumors; pain blockages, and vertebroplasties. Data were analyzed with respect to scanner settings, overall radiation doses, and individual doses of planning CT series, CT intervention, and control CT series.

Results: Eighty-five percent of the total radiation dose was applied during the pre- and post-interventional CT series, leaving only 15% applied by the CT-guided intervention itself. Single slice acquisition was associated with lower doses than continuous CT-fluoroscopy (37 mGy cm vs. 153 mGy cm, p<0.001). The third quartile of radiation doses varied considerably for different interventions. The highest doses were observed in complex interventions like RFA/MWA of the liver, followed by vertebroplasty and RFA/MWA of the lung.

Conclusions: This paper suggests preliminary reference levels for various intervention types and discusses strategies for dose reduction. A multicenter registry of radiation exposure including a broader spectrum of scanners and intervention types is needed to develop definitive reference levels.

Keywords: Image-guided biopsy; Radiation dosage; Radiation protection; Radiology, Interventional; Tomography, Spiral computed.

PubMed Disclaimer