. 2014 Jan-Feb;7(1):92-6.
doi: 10.1016/j.brs.2013.08.004. Epub 2013 Sep 13.

Safety of 5 kHz tACS

Affiliations
Clinical Trial

Safety of 5 kHz tACS

Leila Chaieb et al. Brain Stimul. 2014 Jan-Feb.

Abstract

Background: Sinusoidal transcranial alternating current stimulation (tACS) at 5 kHz applied for 10 min at 1 mA intensity over the hand area of the primary motor cortex (M1) results in sustained changes in cortical excitability as previously demonstrated.

Objective: Here we have assessed safety aspects of this stimulation method by measuring neuron-specific enolase (NSE) levels, examining electroencephalogram (EEG) traces and analyzing anatomical data by using magnetic resonance imaging (MRI).

Methods: Altogether 18 healthy volunteers participated in the study. tACS was applied at 5 kHz for a duration of 10 min over the left M1 at an intensity of 1 mA.

Results: After stimulation no significant changes were detected in NSE levels, no structural alterations were observed in the anatomical scans and no pathological changes were found in the EEG recordings.

Conclusions: Our data imply that the application of tACS is safe at least within these parameters and with these applied protocols.

Keywords: Alternating current stimulation; Human; M1; Side effects.