Comparative Study
. 2013 Dec;133(12):1757-62.
doi: 10.1007/s00402-013-1861-z. Epub 2013 Oct 2.

Early loosening of a press-fit cup with ceramic-on-ceramic articulation: our early results

Affiliations
Comparative Study

Early loosening of a press-fit cup with ceramic-on-ceramic articulation: our early results

Daniël Haverkamp et al. Arch Orthop Trauma Surg. 2013 Dec.

Abstract

Introduction: In this study, we present the short-term results of the Selexys TH+ cup with the Ceramys inlay which is a press-fit cup with a ceramic-on-ceramic articulation. (Mathys, Bettlach, Switzerland). We compared the results with a retrospective-matched control group with a Delta PF cup (Lima, Udine, Italy), which is also a press-fit cup with a ceramic-on-ceramic articulation.

Materials and methods: 257 elective hip arthroplasties with the Selexys TH+ cup in 250 patients placed in 2009 and 2010 were analyzed and compared with a control group retrospective analysis of the uncemented Delta PF cup (Lima, Udine, Italy) placed in 2007 and 2008 in 208 patients (222 hips). Surgical technique and surgeons were identical in both groups.

Results: During a follow-up period of 3-21 months, 19 aseptic loosenings (7.4 %) were found for the Selexys TH+ cup. The survival plotted by a Kaplan-Meier curve shows a 1-year survival of 87.4 %. The Lima Delta PF cup showed a 1-year survival of 99.5 %. Failure analysis showed no clear explanation for this early loosening.

Conclusion: The Selexys TH+ cup combined with the Ceramys ceramic-on-ceramic inlay coupling show an unacceptable high early revision rate. Therefore, we advice against using this combination.