Modular taper junction corrosion and failure: how to approach a recalled total hip arthroplasty implant
- PMID: 24090661
- DOI: 10.1016/j.arth.2013.08.026
Modular taper junction corrosion and failure: how to approach a recalled total hip arthroplasty implant
Abstract
Corrosion at the modular neck-stem taper junction has become an increasingly important topic as several reports have identified this couple as a possible source for early failure with findings similar to failed metal-on-metal hip arthroplasties. Recently, two different modular stem systems from a single manufacturer were voluntarily recalled due to concerns of failure of the modular taper junction. We discuss how to approach the diagnosis and management of patients with these particular stem systems. We further reviewed the literature to evaluate whether this is a manufacturer-specific defect or indicative of a broader trend. Recent studies appear to implicate the basic design of the neck-stem taper junction, rather than a single manufacturer, which is at high risk for fretting and corrosion.
Keywords: THA; corrosion; modular junction corrosion; modular stem.
© 2013.
Comment in
-
Reply to Letter to the editor by Davies et al.J Arthroplasty. 2014 Feb;29(2):450-1. doi: 10.1016/j.arth.2013.10.029. Epub 2013 Nov 14. J Arthroplasty. 2014. PMID: 24332813 No abstract available.
-
Letter to the editor: In response to "Modular taper junction corrosion and failure: how to approach a recalled total hip arthroplasty implant".J Arthroplasty. 2014 Feb;29(2):449-50. doi: 10.1016/j.arth.2013.10.028. Epub 2013 Nov 15. J Arthroplasty. 2014. PMID: 24360493 No abstract available.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical