Low-tilt monophasic and biphasic waveforms compared with standard biphasic waveforms in the transvenous defibrillation of ventricular fibrillation

Affiliations
Comparative Study

Low-tilt monophasic and biphasic waveforms compared with standard biphasic waveforms in the transvenous defibrillation of ventricular fibrillation

Johan R Bennett et al. Pacing Clin Electrophysiol. 2014 Mar.

Abstract

Background: Commercially available implantable defibrillators utilize a high-tilt waveform. Studies in atrial fibrillation and transthoracic defibrillation of ventricular fibrillation (VF) have shown improved defibrillation efficacy using low-tilt (LT) waveforms. We investigated the feasibility, efficacy, and safety of a LT waveform in the transvenous defibrillation of VF and hypothesized that it would be more efficacious than standard tilted biphasic (STB) waveforms.

Methods and results: The investigation was performed in four phases in a porcine model: an efficacy study of LT monophasic waveforms (n = 9), an efficacy study of LT biphasic waveforms (n = 9), a comparison study between the most successful LT waveforms and clinically available STB waveforms (n = 15), and a safety study (n = 9). A total of 1,056 shocks were delivered (phase 1: 288, phase 2: 288, phase 3: 480). The LT biphasic 8/4-ms waveform was significantly more likely to successfully defibrillate than the LT monophasic and STB waveforms with an odds ratio of 122.3 (95% confidence interval: 32.5, 460.2, P < 0.001). The calculated defibrillation threshold (E50) for the LT 8/4-ms waveform was 12.7 J compared to 43.5 J and 45.5 J for STB waveforms 6/6 ms and 8/4 ms, respectively, and 47.7 J for LT 12-ms waveform. The LT 8/4-ms waveform had no lasting detrimental effect on cardiac function, and any transient hemodynamical or biochemical changes observed were comparable to those observed with STB waveforms.

Conclusion: LT waveforms are effective and appear safe in transvenous defibrillation in a porcine model of VF. The LT biphasic 8/4-ms waveform is more efficacious than conventional waveforms.

Keywords: defibrillation; low tilt; ventricular fibrillation; waveform.