Validation of cardiac output using real-time measurement of oxygen consumption during cardiac catheterization in children under 3 years of age
- PMID: 24102795
- DOI: 10.1111/chd.12140
Validation of cardiac output using real-time measurement of oxygen consumption during cardiac catheterization in children under 3 years of age
Abstract
Objective: To validate a method for determination of cardiac index (CI) using real-time measurement of oxygen consumption (VO2 ) in young children undergoing cardiac catheterization.
Design: Retrospective review comparing thermodilution cardiac index (TDCI) to CI calculated by the Fick equation using real-time measured VO2 (RT-VO2 ) and VO2 derived from 2 published predictive equations. Paired t-test and Bland-Altman analysis were used to compare TDCI to Fick CI. A survey to ascertain pediatric cardiac catheterization practices regarding VO2 determination was also conducted.
Setting: Quaternary care children's hospital cardiac catheterization laboratory.
Patients: Children <3 years old with structurally normal hearts undergoing cardiac catheterization under general anesthesia with at least one set of contemporaneous TDCI and RT-VO2 measurements.
Results: Thirty-six paired measurements of TDCI and RT-VO2 were made in 27 patients over a 2-year period. Indications for catheterization included congenital diaphragmatic hernia postrepair (n = 13), heart disease post-orthotopic heart transplant (n = 13), and suspected cardiomyopathy (n = 1). Mean age was 21.5 ± 8 months; median weight was 9.9 kg (IQR 8.57, 12.2). RT-VO2 was higher than VO2 predicted by the LaFarge equation (190 ± 31 vs. 173.8 ± 12.8 mL/min/m(2), P < .001), but there was no difference between TDCI and Fick CI calculated using VO2 from any method. Bland-Altman analysis showed excellent agreement between TDCI and Fick CI using RT-VO2 and VO2 predicted by the Lundell equation; Fick CI using VO2 predicted by the LaFarge equation showed fair agreement with TDCI.
Conclusions: In children <3 years with a structurally normal heart, RT-VO2 generates highly accurate determinations of Fick CI as compared with TDCI. Additionally, in this population, VO2 derived from the LaFarge and Lundell equations generates accurate Fick CI compared with TDCI. Future studies are needed to identify factors associated with inaccurate VO2 generated from these predictive equations.
Keywords: Cardiac Catheterization; Cardiac Output; Hemodynamics; Oxygen Consumption; Pediatric Cardiology.
© 2013 Wiley Periodicals, Inc.
Similar articles
-
The Estimated Fick Method Systematically Over-Estimates Cardiac Index Compared to Thermodilution in Children.Pediatr Cardiol. 2025 Mar;46(3):713-718. doi: 10.1007/s00246-024-03491-z. Epub 2024 May 18. Pediatr Cardiol. 2025. PMID: 38762577
-
Effect of solute temperature in the measurement of cardiac output in children using the thermodilution technique.Congenit Heart Dis. 2017 Mar;12(2):181-187. doi: 10.1111/chd.12423. Epub 2016 Nov 4. Congenit Heart Dis. 2017. PMID: 27813342
-
Real-time gas-exchange measurement of oxygen consumption in neonates and infants after cardiac surgery.Crit Care Med. 1993 Sep;21(9):1369-75. doi: 10.1097/00003246-199309000-00022. Crit Care Med. 1993. PMID: 8370302
-
Accurate measurement of oxygen consumption in children undergoing cardiac catheterization.Catheter Cardiovasc Interv. 2013 Jan 1;81(1):125-32. doi: 10.1002/ccd.24440. Epub 2012 Nov 8. Catheter Cardiovasc Interv. 2013. PMID: 22488802 Review.
-
Continuous measurement of cardiac output by the Fick principle in infants and children: comparison with the thermodilution method.Intensive Care Med. 1996 May;22(5):467-71. doi: 10.1007/BF01712169. Intensive Care Med. 1996. PMID: 8796404 Review.
Cited by
-
Utility of Echocardiography in the Assessment of Left Ventricular Diastolic Function and Restrictive Physiology in Children and Young Adults with Restrictive Cardiomyopathy: A Comparative Echocardiography-Catheterization Study.Pediatr Cardiol. 2017 Feb;38(2):381-389. doi: 10.1007/s00246-016-1526-0. Epub 2016 Nov 23. Pediatr Cardiol. 2017. PMID: 27878632 Free PMC article.
-
Measurement of Cardiac Output Using an Ultrasonic Cardiac Output Monitor (USCOM) in Patients with Single-Ventricle Physiology.Pediatr Cardiol. 2022 Aug;43(6):1205-1213. doi: 10.1007/s00246-022-02840-0. Epub 2022 Feb 6. Pediatr Cardiol. 2022. PMID: 35124709
-
The Estimated Fick Method Systematically Over-Estimates Cardiac Index Compared to Thermodilution in Children.Pediatr Cardiol. 2025 Mar;46(3):713-718. doi: 10.1007/s00246-024-03491-z. Epub 2024 May 18. Pediatr Cardiol. 2025. PMID: 38762577
-
Measured Oxygen Consumption During Pediatric Cardiac Catheterization is More Accurate than Assumed Oxygen Consumption.Pediatr Cardiol. 2024 Oct;45(7):1466-1474. doi: 10.1007/s00246-023-03186-x. Epub 2023 May 27. Pediatr Cardiol. 2024. PMID: 37243747
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources