Noise-tolerant instantaneous heart rate and R-peak detection using short-term autocorrelation for wearable healthcare systems
- PMID: 24111438
- DOI: 10.1109/EMBC.2013.6611251
Noise-tolerant instantaneous heart rate and R-peak detection using short-term autocorrelation for wearable healthcare systems
Abstract
This paper describes a robust method of Instantaneous Heart Rate (IHR) and R-peak detection from noisy electrocardiogram (ECG) signals. Generally, the IHR is calculated from the R-wave interval. Then, the R-waves are extracted from the ECG using a threshold. However, in wearable bio-signal monitoring systems, noise increases the incidence of misdetection and false detection of R-peaks. To prevent incorrect detection, we introduce a short-term autocorrelation (STAC) technique and a small-window autocorrelation (SWAC) technique, which leverages the similarity of QRS complex waveforms. Simulation results show that the proposed method improves the noise tolerance of R-peak detection.