. 2013 Sep;10(3):253-7.
doi: 10.3969/j.issn.1671-5411.2013.03.012.

OptiVol fluid index predicts acute decompensation of heart failure with a high rate of unexplained events

Affiliations

OptiVol fluid index predicts acute decompensation of heart failure with a high rate of unexplained events

Xin-Wei Yang et al. J Geriatr Cardiol. 2013 Sep.

Abstract

Background: Intrathoracic impedance monitoring has emerged as a promising new technique for the detection of impending heart failure (HF). Although false positive episodes have been reported in case reports and clinical trials, the efficacy and false positive rate in real-world practice remain unclear.

Objective: The aim of this study is to investigate the utility and reliability of the OptiVol alert feature in clinical practice.

Methods: We continuously recruited patients who underwent implantable cardioverter-defibrillator (ICD) or cardiac resynchronization therapy with defibrillator (CRT-D) implantation with feature of intrathoracic impedance monitoring system in our center from Sep. 2010 to Oct. 2012. Regular in-office follow-up were required of all patients and the following information was collected at each visit: medical history, device interrogation, N-terminal pro-brain natriuretic peptide (NT-proBNP) measurement and an echocardiogram. Worsening HF was defined as hospitalization or the presentation of signs or symptoms of HF.

Results: FORTY THREE PATIENTS (MALE: 76.7%, mean age: 57 ± 15 years, left ventricular ejection fraction (LVEF): 33% ± 14%) were included in this observational study. Fifty four alert events and 14 adjudicated worsening HF were detected within 288 ±163 days follow-up. Eleven (20.4%) alert episodes were associated with acute cardiac decompensation in 9 patients with a positive predictive value of 78.6%. Forty three audible alerts showed no connection to worsening HF. The unexplained alerts rate was 79.6% and 1.27 per person-year. Thirty seven alarm alerts were detected in patients with EF < 45%, among which 9 accompanied with HF, 17 alerts detected in patients with LVEF ≥ 45% and 2 associated with HF. There was no significant difference between the two groups (9/37 vs. 2/17; P = 0.47).

Conclusions: Patients with normal or nearly normal left ventricular systolic function also exhibited considerable alert events. The OptiVol fluid index predicted worsening cardiac events with a high unexplained detection rate, and any alert must therefore be analyzed with great caution. Efforts to improve the specificity of this monitoring system represent a significant aspect of future studies.

Keywords: Heart failure; Intrathoracic impedance measurement; Left ventricular ejection fraction; OptiVol fluid index.

PubMed Disclaimer

Figures

Figure 1.
Figure 1.. Thoracic impedance and OptiVol FI trend recording.
OptiVol fluid index is an accumulation of the difference between the daily and reference impedance.
Figure 2.
Figure 2.. Incidence of alert events in relation to heart failure in different ejection fraction (EF) groups.

Similar articles

Cited by

References

    1. Roger VL, Go AS, Lloyd-Jones DM, et al. et al. Heart disease and stroke statistics--2012 update: a report from the American Heart Association. Circulation. 2012;125:E2–E220. - PMC - PubMed
    1. Wang L, Lahtinen S, Lentz L, et al. et al. Feasibility of using an implantable system to measure thoracic congestion in an ambulatory chronic heart failure canine model. Pacing Clin Electrophysiol. 2005;28:404–411. - PubMed
    1. Ganion V, Rhodes M, Stadler RW. Intrathoracic impedance to monitor heart failure status: a comparison of two methods in a chronic heart failure dog model. Congest Heart Fail. 2005;11:177–181, 211. - PubMed
    1. Becher J, Kaufmann SG, Paule S, et al. et al. Device-based impedance measurement is a useful and accurate tool for direct assessment of intrathoracic fluid accumulation in heart failure. Europace. 2010;12:731–740. - PubMed
    1. Yu CM, Wang L, Chau E, et al. et al. Intrathoracic impedance monitoring in patients with heart failure: correlation with fluid status and feasibility of early warning preceding hospitalization. Circulation. 2005;112:841–848. - PubMed

LinkOut - more resources