. 2013 Oct;7(5):601-9.
doi: 10.1109/TBCAS.2013.2281772. Epub 2013 Oct 17.

Massively-parallel neuromonitoring and neurostimulation rodent headset with nanotextured flexible microelectrodes

Massively-parallel neuromonitoring and neurostimulation rodent headset with nanotextured flexible microelectrodes

Arezu Bagheri et al. IEEE Trans Biomed Circuits Syst. 2013 Oct.

Abstract

We present a compact wireless headset for simultaneous multi-site neuromonitoring and neurostimulation in the rodent brain. The system comprises flexible-shaft microelectrodes, neural amplifiers, neurostimulators, a digital time-division multiplexer (TDM), a micro-controller and a ZigBee wireless transceiver. The system is built by parallelizing up to four 0.35 μm CMOS integrated circuits (each having 256 neural amplifiers and 64 neurostimulators) to provide a total maximum of 1024 neural amplifiers and 256 neurostimulators. Each bipolar neural amplifier features 54 dB-72 dB adjustable gain, 1 Hz-5 kHz adjustable bandwidth with an input-referred noise of 7.99 μVrms and dissipates 12.9 μW. Each current-mode bipolar neurostimulator generates programmable arbitrary-waveform biphasic current in the range of 20-250 μA and dissipates 2.6 μW in the stand-by mode. Reconfigurability is provided by stacking a set of dedicated mini-PCBs that share a common signaling bus within as small as 22 × 30 × 15 mm³ volume. The system features flexible polyimide-based microelectrode array design that is not brittle and increases pad packing density. Pad nanotexturing by electrodeposition reduces the electrode-tissue interface impedance from an average of 2 MΩ to 30 kΩ at 100 Hz. The rodent headset and the microelectrode array have been experimentally validated in vivo in freely moving rats for two months. We demonstrate 92.8 percent seizure rate reduction by responsive neurostimulation in an acute epilepsy rat model.