. 2013 Oct 18;4(4):161-6.
doi: 10.5312/wjo.v4.i4.161. eCollection 2013.

Trunnionosis: A pain in the neck

Affiliations

Trunnionosis: A pain in the neck

Philip S Pastides et al. World J Orthop. .

Abstract

Metal-on-metal (MoM) hip replacements have proven to be a modern day orthopaedic failure. The early enthusiasm and promise of a hard, durable bearing was quickly quashed following the unanticipated wear rates. The release of metal ions into the blood stream has been shown to lead to surrounding soft tissue complications and early failure. The devastating destruction caused has led to a large number of revision procedures and implant extractions. The resulting research into this field has led to a new area of interest; that of the wear at the trunnion of the prosthesis. It had been previously thought that the metal debris was generated solely from the weight bearing articulation, however with the evolution of modularity to aid surgical options, wear at the trunnion is becoming more apparent. The phenomenon of "trunnionosis" is a rapidly developing area of interest that may contribute to the overall effect of metallosis in MoM replacements but may also lead to the release of metal ions in non MoM hip designs. The aim of this paper is to introduce, explain and summarise the evidence so far in the field of trunnionosis. The evidence for this phenomenon, the type of debris particles generated and a contrast between MoM, non MoM and resurfacing procedures are also presented.

Keywords: Arthroplasty; Hip; Ions; Trunnion; Wear.

PubMed Disclaimer

References

    1. Mann BS, Whittingham-Jones PM, Shaerf DA, Nawaz ZS, Harvie P, Hart AJ, Skinner JA. Metal-on-metal bearings, inflammatory pseudotumours and their neurological manifestations. Hip Int. 2012;22:129–136. - PubMed
    1. Haddad FS, Thakrar RR, Hart AJ, Skinner JA, Nargol AV, Nolan JF, Gill HS, Murray DW, Blom AW, Case CP. Metal-on-metal bearings: the evidence so far. J Bone Joint Surg Br. 2011;93:572–579. - PubMed
    1. Lombardi AV, Barrack RL, Berend KR, Cuckler JM, Jacobs JJ, Mont MA, Schmalzried TP. The Hip Society: algorithmic approach to diagnosis and management of metal-on-metal arthroplasty. J Bone Joint Surg Br. 2012;94:14–18. - PubMed
    1. Langton DJ, Sidaginamale RP, Joyce TJ, Natu S, Blain P, Jefferson RD, Rushton S, Nargol AV. The clinical implications of elevated blood metal ion concentrations in asymptomatic patients with MoM hip resurfacings: a cohort study. BMJ Open. 2013;3:pii: e001541. - PMC - PubMed
    1. Archibeck MJ, Cummins T, Carothers J, Junick DW, White RE. A comparison of two implant systems in restoration of hip geometry in arthroplasty. Clin Orthop Relat Res. 2011;469:443–446. - PMC - PubMed