. 2012 Sep 1;11(3):409-17.
eCollection 2012.

Bioharness(™) Multivariable Monitoring Device: Part. II: Reliability

Affiliations

Bioharness(™) Multivariable Monitoring Device: Part. II: Reliability

James A Johnstone et al. J Sports Sci Med. .

Abstract

The Bioharness(™) monitoring system may provide physiological information on human performance but the reliability of this data is fundamental for confidence in the equipment being used. The objective of this study was to assess the reliability of each of the 5 Bioharness(™) variables using a treadmill based protocol. 10 healthy males participated. A between and within subject design to assess the reliability of Heart rate (HR), Breathing Frequency (BF), Accelerometry (ACC) and Infra-red skin temperature (ST) was completed via a repeated, discontinuous, incremental treadmill protocol. Posture (P) was assessed by a tilt table, moved through 160°. Between subject data reported low Coefficient of Variation (CV) and strong correlations(r) for ACC and P (CV< 7.6; r = 0.99, p < 0.01). In contrast, HR and BF (CV~19.4; r~0.70, p < 0.01) and ST (CV 3.7; r = 0.61, p < 0.01), present more variable data. Intra and inter device data presented strong relationships (r > 0.89, p < 0.01) and low CV (<10.1) for HR, ACC, P and ST. BF produced weaker relationships (r < 0.72) and higher CV (<17.4). In comparison to the other variables BF variable consistently presents less reliability. Global results suggest that the Bioharness(™) is a reliable multivariable monitoring device during laboratory testing within the limits presented. Key pointsHeart rate and breathing frequency data increased in variance at higher velocities (i.e. ≥ 10 km.h(-1))In comparison to the between subject testing, the intra and inter reliability presented good reliability in data suggesting placement or position of device relative to performer could be important for data collectionUnderstanding a devices variability in measurement is important before it can be used within an exercise testing or monitoring setting.

Keywords: Physiological technology; exercise; reproducibility of measurement.

PubMed Disclaimer

Figures

Figure 1.
Figure 1.
Residual versus predicted plot demonstrating the data spread for (a) HR and (b) BF.

References

    1. Achten J., Jeukendrup E.(2003) Heart rate monitoring: applications and limitations. Sports Medicine 333(7), 517–538 - PubMed
    1. Atkinson G., Nevill A.M.(1998) Statistical methods for assessing measurement error (reliability) in variables relevant to sports medicine. Sports Medicine 226(4), 217–238 - PubMed
    1. Bernmark E., Wiktorin C.(2002) A triaxial accelerometer for measuring arm movements. Applied Ergonomics 333, 541–547 - PubMed
    1. Boudet G., Chamoux A.(2000) Heart rate monitors and abnormal heart rhythm detection. Archives of Physiology and Biochemistry 1108(4), 371–379 - PubMed
    1. Brage S., Brage N., Franks P.W., Ekelund U., Wareham N.J.(2005) Reliability and validity of the combined heart rate and movement sensor Actiheart. European Journal of Clinical Nutrition 559, 561–570 - PubMed

LinkOut - more resources