Field based reliability and validity of the bioharness™ multivariable monitoring device
- PMID: 24150074
- PMCID: PMC3763310
Field based reliability and validity of the bioharness™ multivariable monitoring device
Abstract
The Bioharness™ device is designed for monitoring physiological variables in free-living situations but has only been proven to be reliable and valid in a laboratory environment. Therefore, this study aimed to determine the reliability and validity of the Bioharness™ using a field based protocol. Twenty healthy males participated. Heart rate (HR), breathing frequency (BF) and accelerometry (ACC) were assessed by simultaneous measurement of two Bioharness™ devices and a test-retest of a discontinuous incremental walk-jog-run protocol (4 - 11 km·h(-1)) completed in a sports hall. Adopted precision of measurement devices were; HR: Polar T31 (Polar Electro), BF: Spirometer (Cortex Metalyser), ACC: Oxygen expenditure (Cortex Metalyser). For all data, precision of measurement reported good relationships (r = 0.61 to 0.67, p < 0.01) and large Limits of Agreement for HR (>79.2 b·min(-1)) and BF (>54.7 br·min(-1)). ACC presented excellent precision (r = 0.94, p < 0.01). Results for HR (r= ~0.91, p < 0.01: CV <7.6) and ACC (r > 0.97, p < 0.01; CV <14.7) suggested these variables are reliable. BF presented more variable data (r = 0.46-0.61, p < 0.01; CV < 23.7). As velocity of movement increased (>8 km·h(-1)) data became more erroneous. A data cleaning protocol removed gross errors in the data analysis and subsequent reliability and validity statistics improved across all variables. In conclusion, the Bioharness™ HR and ACC variables have demonstrated reliability and validity in a field setting, though data collected at higher velocities should be treated with caution. Measuring human physiological responses in a field based environment allows for more ecologically valid data to be collected and devices such as the Bioharness™ could be used by exercise professionals to begin to further investigate this area.
Keywords: Multi-variable; ecological validity; new technology; physiological monitoring.
Figures
References
-
- Achten J., Jeukendrup E.(2003) Heart rate monitoring: Applications and limitations. Journal of Sports Medicine 33, 517-538 - PubMed
-
- Astrand P., Rohdahl K., Dahl H., Stromme S.(2003) Textbook of work physiology, Champaign IL, Human Kinetics.
-
- Atkinson G., Nevill A.M.(1998) Statistical methods for assessing measurement error (reliability) in variables relevant to sports medicine. Sports Medicine 26, 217-238 - PubMed
-
- Bartlett R.(2006) Medicine and science in cricket. Journal of Science and Medicine in Sport 9, 470-471 - PubMed
-
- Bartlett R., Wheat J., Robins M.(2007)Is movement variability important for sports biomechanists? Sports Biomechanics 6, 223-243 - PubMed