Abstract
With growing complexity in radiotherapy treatment delivery, it has become mandatory to check each and every treatment plan before implementing clinically. This process is currently administered by an independent secondary check of all treatment parameters and as a pre-treatment quality assurance (QA) check for intensity modulated radiation therapy (IMRT) and volumetric modulated arc therapy treatment plans. Although pre-treatment IMRT QA is aimed to ensure the correct dose is delivered to the patient, it does not necessarily predict the clinically relevant patient dose errors. During radiotherapy, treatment uncertainties can affect tumor control and may increase complications to surrounding normal tissues. To combat this, image guided radiotherapy is employed to help ensure the plan conditions are mimicked on the treatment machine. However, it does not provide information on actual delivered dose to the tumor volume. Knowledge of actual dose delivered during treatment aid in confirming the prescribed dose and also to replan/reassess the treatment in situations where the planned dose is not delivered as expected by the treating physician. Major accidents in radiotherapy would have been averted if real time dosimetry is incorporated as part of the routine radiotherapy procedure. Of late real-time dosimetry is becoming popular with complex treatments in radiotherapy. Real-time dosimetry can be either in the form of point doses or planar doses or projected on to a 3D image dataset to obtain volumetric dose. They either provide entrance dose or exit dose or dose inside the natural cavities of a patient. In external beam radiotherapy, there are four different established platforms whereby the delivered dose information can be obtained: (1) Collimator; (2) Patient; (3) Couch; and (4) Electronic Portal Imaging Device. Current real-time dosimetric techniques available in radiotherapy have their own advantages and disadvantages and a combination of one or more of these methods provide vital information about the actual dose delivered to radiotherapy patients.
Keywords: Cancer; Dosimetry; External beam; Radiotherapy; Real-time.
Similar articles
-
A feasibility study of using couch-based real time dosimetric device in external beam radiotherapy.Med Phys. 2011 Dec;38(12):6539-52. doi: 10.1118/1.3660773. Med Phys. 2011. PMID: 22149836
-
Dosimetric quality, accuracy, and deliverability of modulated radiotherapy treatments for spinal metastases.Med Dosim. 2016 Autumn;41(3):258-66. doi: 10.1016/j.meddos.2016.06.006. Med Dosim. 2016. PMID: 27545010
-
A novel approach to SBRT patient quality assurance using EPID-based real-time transit dosimetry : A step to QA with in vivo EPID dosimetry.Strahlenther Onkol. 2020 Feb;196(2):182-192. doi: 10.1007/s00066-019-01549-z. Epub 2020 Jan 10. Strahlenther Onkol. 2020. PMID: 31925465
-
A literature review of electronic portal imaging for radiotherapy dosimetry.Radiother Oncol. 2008 Sep;88(3):289-309. doi: 10.1016/j.radonc.2008.07.008. Epub 2008 Aug 14. Radiother Oncol. 2008. PMID: 18706727 Review.
-
Dose to normal tissues outside the radiation therapy patient's treated volume: a review of different radiation therapy techniques.Health Phys. 2008 Nov;95(5):666-76. doi: 10.1097/01.HP.0000326342.47348.06. Health Phys. 2008. PMID: 18849701 Review.
Cited by
-
Hydrogenated Amorphous Silicon Charge-Selective Contact Devices on a Polyimide Flexible Substrate for Dosimetry and Beam Flux Measurements.Sensors (Basel). 2025 Feb 19;25(4):1263. doi: 10.3390/s25041263. Sensors (Basel). 2025. PMID: 40006492 Free PMC article.
References
-
- Kuban D, Pollack A, Huang E, Levy L, Dong L, Starkschall G, Rosen I. Hazards of dose escalation in prostate cancer radiotherapy. Int J Radiat Oncol Biol Phys. 2003;57:1260–1268. - PubMed
-
- Hurkmans CW, Cho BC, Damen E, Zijp L, Mijnheer BJ. Reduction of cardiac and lung complication probabilities after breast irradiation using conformal radiotherapy with or without intensity modulation. Radiother Oncol. 2002;62:163–171. - PubMed
-
- Tubiana M, Eschwège F. Conformal radiotherapy and intensity-modulated radiotherapy--clinical data. Acta Oncol. 2000;39:555–567. - PubMed
-
- Vatnitsky S, Ortiz Lopez P, Izewska J, Meghzifene A, Levin V. The radiation overexposure of radiotherapy patients in Panama 15 June 2001. Radiother Oncol. 2001;60:237–238. - PubMed