Low-dose CT coronary angiography using iterative reconstruction with a 256-slice CT scanner
- PMID: 24198908
- PMCID: PMC3817280
- DOI: 10.4330/wjc.v5.i10.382
Low-dose CT coronary angiography using iterative reconstruction with a 256-slice CT scanner
Abstract
Aim: To explore whether computer tomography coronary angiography (CTCA) using iterative reconstruction (IR) leads to significant radiation dose reduction without a significant loss in image interpretability compared to conventional filtered back projection (FBP).
Methods: A consecutive series of 200 patients referred to our institution to undergo CTCA constituted the study population. Patients were sequentially assigned to FBP or IR. All studies were acquired with a 256-slice CT scanner. A coronary segment was considered interpretable if image quality was adequate for evaluation of coronary lesions in all segments ≥ 1.5 mm.
Results: The mean age was 56.3 ± 9.6 years and 165 (83%) were male, with no significant differences between groups. Most scans were acquired using prospective ECG triggering, without differences between groups (FBP 84% vs IR 82%; P = 0.71). A total of 3198 (94%) coronary segments were deemed of diagnostic quality. The percent assessable coronary segments was similar between groups (FBP 91.7% ± 4.0% vs IR 92.5% ± 2.8%; P = 0.12). Radiation dose was significantly lower in the IR group (2.8 ± 1.4 mSv vs 4.6 ± 3.0 mSv; P < 0.0001). Image noise (37.8 ± 1.4 HU vs 38.2 ± 2.4 HU; P = 0.20) and signal density (461.7 ± 51.9 HU vs 462.2 ± 51.2 HU; P = 0.54) levels did not differ between FBP and IR groups, respectively. The IR group was associated to significant effective dose reductions, irrespective of the acquisition mode.
Conclusion: Application of IR in CTCA preserves image interpretability despite a significant reduction in radiation dose.
Keywords: Iterative reconstruction; Low-dose computer tomography coronary angiography.
Similar articles
-
Iterative reconstruction to preserve image quality and diagnostic accuracy at reduced radiation dose in coronary CT angiography: an intraindividual comparison.JACC Cardiovasc Imaging. 2013 Dec;6(12):1239-49. doi: 10.1016/j.jcmg.2013.08.008. Epub 2013 Oct 23. JACC Cardiovasc Imaging. 2013. PMID: 24269265
-
Iterative model reconstruction: improved image quality of low-tube-voltage prospective ECG-gated coronary CT angiography images at 256-slice CT.Eur J Radiol. 2014 Aug;83(8):1408-15. doi: 10.1016/j.ejrad.2014.04.027. Epub 2014 May 2. Eur J Radiol. 2014. PMID: 24873832
-
Prospectively ECG-triggered high-pitch coronary angiography with third-generation dual-source CT at 70 kVp tube voltage: feasibility, image quality, radiation dose, and effect of iterative reconstruction.J Cardiovasc Comput Tomogr. 2014 Nov-Dec;8(6):418-25. doi: 10.1016/j.jcct.2014.09.003. Epub 2014 Sep 16. J Cardiovasc Comput Tomogr. 2014. PMID: 25439789
-
Accuracy of prospectively ECG-triggered very low-dose coronary dual-source CT angiography using iterative reconstruction for the detection of coronary artery stenosis: comparison with invasive catheterization.Eur Heart J Cardiovasc Imaging. 2014 Nov;15(11):1238-45. doi: 10.1093/ehjci/jeu113. Epub 2014 Jun 17. Eur Heart J Cardiovasc Imaging. 2014. PMID: 24939952
-
Intra-individual diagnostic image quality and organ-specific-radiation dose comparison between spiral cCT with iterative image reconstruction and z-axis automated tube current modulation and sequential cCT.Eur J Radiol Open. 2016 Jul 26;3:182-90. doi: 10.1016/j.ejro.2016.05.006. eCollection 2016. Eur J Radiol Open. 2016. PMID: 27504476 Free PMC article. Review.
Cited by
-
128-slice CT angiography of the aorta without ECG-gating: efficacy of faster gantry rotation time and iterative reconstruction in terms of image quality and radiation dose.Eur Radiol. 2016 Feb;26(2):359-69. doi: 10.1007/s00330-015-3848-3. Epub 2015 May 24. Eur Radiol. 2016. PMID: 26003792
-
Dose reduction with iterative reconstruction for coronary CT angiography: a systematic review and meta-analysis.Br J Radiol. 2016;89(1058):20150068. doi: 10.1259/bjr.20150068. Epub 2015 Nov 12. Br J Radiol. 2016. PMID: 26562096 Free PMC article.
-
Development of an organ-specific insert phantom generated using a 3D printer for investigations of cardiac computed tomography protocols.J Med Radiat Sci. 2018 Sep;65(3):175-183. doi: 10.1002/jmrs.279. Epub 2018 Apr 30. J Med Radiat Sci. 2018. PMID: 29707915 Free PMC article.
-
The role of advanced reconstruction algorithms in cardiac CT.Cardiovasc Diagn Ther. 2017 Oct;7(5):527-538. doi: 10.21037/cdt.2017.08.12. Cardiovasc Diagn Ther. 2017. PMID: 29255694 Free PMC article. Review.
-
Is there a role for a local inpatient CT coronary angiography service in selected patients with acute coronary syndrome? A cohort analysis of inpatient tertiary centre referrals for invasive coronary angiography.Open Heart. 2016 Mar 10;3(1):e000389. doi: 10.1136/openhrt-2015-000389. eCollection 2016. Open Heart. 2016. PMID: 27042324 Free PMC article.
References
-
- Hendel RC, Patel MR, Kramer CM, Poon M, Hendel RC, Carr JC, Gerstad NA, Gillam LD, Hodgson JM, Kim RJ, et al. ACCF/ACR/SCCT/SCMR/ASNC/NASCI/SCAI/SIR 2006 appropriateness criteria for cardiac computed tomography and cardiac magnetic resonance imaging: a report of the American College of Cardiology Foundation Quality Strategic Directions Committee Appropriateness Criteria Working Group, American College of Radiology, Society of Cardiovascular Computed Tomography, Society for Cardiovascular Magnetic Resonance, American Society of Nuclear Cardiology, North American Society for Cardiac Imaging, Society for Cardiovascular Angiography and Interventions, and Society of Interventional Radiology. J Am Coll Cardiol. 2006;48:1475–1497. - PubMed
-
- Greenland P, Bonow RO, Brundage BH, Budoff MJ, Eisenberg MJ, Grundy SM, Lauer MS, Post WS, Raggi P, Redberg RF, et al. ACCF/AHA 2007 clinical expert consensus document on coronary artery calcium scoring by computed tomography in global cardiovascular risk assessment and in evaluation of patients with chest pain: a report of the American College of Cardiology Foundation Clinical Expert Consensus Task Force (ACCF/AHA Writing Committee to Update the 2000 Expert Consensus Document on Electron Beam Computed Tomography) developed in collaboration with the Society of Atherosclerosis Imaging and Prevention and the Society of Cardiovascular Computed Tomography. J Am Coll Cardiol. 2007;49:378–402. - PubMed
-
- Taylor AJ, Cerqueira M, Hodgson JM, Mark D, Min J, O’Gara P, Rubin GD, Kramer CM, Berman D, Brown A, et al. ACCF/SCCT/ACR/AHA/ASE/ASNC/NASCI/SCAI/SCMR 2010 appropriate use criteria for cardiac computed tomography. A report of the American College of Cardiology Foundation Appropriate Use Criteria Task Force, the Society of Cardiovascular Computed Tomography, the American College of Radiology, the American Heart Association, the American Society of Echocardiography, the American Society of Nuclear Cardiology, the North American Society for Cardiovascular Imaging, the Society for Cardiovascular Angiography and Interventions, and the Society for Cardiovascular Magnetic Resonance. J Am Coll Cardiol. 2010;56:1864–1894. - PubMed
-
- Bischoff B, Hein F, Meyer T, Krebs M, Hadamitzky M, Martinoff S, Schömig A, Hausleiter J. Comparison of sequential and helical scanning for radiation dose and image quality: results of the Prospective Multicenter Study on Radiation Dose Estimates of Cardiac CT Angiography (PROTECTION) I Study. AJR Am J Roentgenol. 2010;194:1495–1499. - PubMed
-
- Hausleiter J, Martinoff S, Hadamitzky M, Martuscelli E, Pschierer I, Feuchtner GM, Catalán-Sanz P, Czermak B, Meyer TS, Hein F, et al. Image quality and radiation exposure with a low tube voltage protocol for coronary CT angiography results of the PROTECTION II Trial. JACC Cardiovasc Imaging. 2010;3:1113–1123. - PubMed
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous