Analytical performance of QMS everolimus assay on ortho Vitros 5,1 FS fusion analyzer: measuring everolimus trough levels for solid organ transplant recipients
- PMID: 24232125
- DOI: 10.1097/FTD.0b013e3182a3b3f6
Analytical performance of QMS everolimus assay on ortho Vitros 5,1 FS fusion analyzer: measuring everolimus trough levels for solid organ transplant recipients
Abstract
Background: Everolimus has recently been approved by Food and Drug Administration for graft maintenance in liver transplant recipients. This drug has a narrow therapeutic index and benefits from close blood level monitoring. Currently, in the United States, the Thermo Fisher Scientific Quantitative Microsphere System (QMS) Everolimus Immunoassay is the only Food and Drug Administration-cleared immunoassay for monitoring everolimus in renal transplant recipients. However, studies on this assay adapted to the Ortho Vitros 5,1 FS chemistry analyzer have not been published, and data of this assay applied to monitoring drug levels in liver transplant recipients are limited. Here, the authors evaluated and validated the QMS everolimus assay on the Vitros analyzer and its application to supporting the immunosuppressant management of mainly liver transplant recipients.
Methods: The analysis was performed according to the QMS assay package insert. The method was compared with a liquid chromatography-tandem mass spectrometry method from a reference laboratory using a total of 34 samples from 1 double lung and liver, 8 liver, and 3 kidney recipients. The method comparison was assessed by Deming regression. Proficiency test materials issued by Everolimus TDM Proficiency Support Program were tested and compared with the peer group results of using the QMS kits.
Results: The assay was linear in the range of 0.75-20.0 ng/mL. Limit of detection was 0.70 ng/mL and lower limit of quantitation was 0.75 ng/mL. Within-day and between-day (20 days) coefficients of variation were between 3.1% and 16.5% at mean levels of 5.3, 12.0, and 17.2 ng/mL, respectively. We obtained a Deming regression of y = 1.271 - 0.666 (r = 0.880) when comparing with the liquid chromatography-tandem mass spectrometry method.
Conclusions: The authors concluded that the analytical performance of the QMS everolimus immunoassay by the Vitros 5,1 FS analyzer was satisfactory for monitoring drug levels of solid organ transplant patients.
Similar articles
-
Evaluation of QMS everolimus assay using Hitachi 917 Analyzer: comparison with liquid chromatography/mass spectrometry.Ther Drug Monit. 2011 Apr;33(2):149-54. doi: 10.1097/FTD.0b013e31820afc97. Ther Drug Monit. 2011. PMID: 21266941
-
Evaluation of QMS everolimus assay using Indiko analyzer: comparison with an ultra-performance liquid chromatography-tandem mass spectrometry method.Ther Drug Monit. 2015 Apr;37(2):275-8. doi: 10.1097/FTD.0000000000000133. Ther Drug Monit. 2015. PMID: 25254414
-
Comparison of everolimus QMS immunoassay on Architect ci4100 and liquid chromatography/mass spectrometry: lack of agreement in organ-transplanted patients.Ther Drug Monit. 2015 Apr;37(2):214-9. doi: 10.1097/FTD.0000000000000126. Ther Drug Monit. 2015. PMID: 25162218
-
The role of therapeutic monitoring of everolimus in solid organ transplantation.Ther Drug Monit. 2005 Oct;27(5):666-76. doi: 10.1097/01.ftd.0000175911.70172.2e. Ther Drug Monit. 2005. PMID: 16175142 Review.
-
Recent advances in analytical methods for the therapeutic drug monitoring of immunosuppressive drugs.Drug Test Anal. 2018 Jan;10(1):81-94. doi: 10.1002/dta.2290. Epub 2017 Sep 28. Drug Test Anal. 2018. PMID: 28851030 Review.
Cited by
-
Effectiveness of central venous pressure versus stroke volume variation in guiding fluid management in renal transplantation.Am J Transl Res. 2021 Jul 15;13(7):7848-7856. eCollection 2021. Am J Transl Res. 2021. PMID: 34377262 Free PMC article.
-
Long-term cross-validation of everolimus therapeutic drug monitoring assays: the Zortracker study.Ther Drug Monit. 2015 Jun;37(3):296-303. doi: 10.1097/FTD.0000000000000191. Ther Drug Monit. 2015. PMID: 25970506 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical