. 1992 May;3(4):398-408.
doi: 10.1016/1044-0305(92)87067-9.

Collisional focusing effects in radio frequency quadrupoles

Affiliations

Collisional focusing effects in radio frequency quadrupoles

D J Douglas et al. J Am Soc Mass Spectrom. 1992 May.

Abstract

The transmission of ions through a conventional two-dimensional radiofrequency-only (rf) quadrupole has been studied for comparatively high operating pressures between 5 × 10(-4) and 1 × 10(-2) torr. Measurements of signals from mass-resolved analyte ions and total ion currents show that, provided the initial injection ion energy is low (1-30 eV), the ion transmission observed through a small aperture at the exit of the rf quadrupole first increases as the gas pressure increases, reaching a maximum at - 8 × 10(-3) torr before decreasing at higher pressures. This is in direct contrast to the expectations of classical scattering. This "collisional focusing" appears to be analogous to effects seen in three-dimensional ion traps. The collisional focusing increases with the mass of the ion (not mass-to-charge ratio) for masses up to at least 16,950 u. The collisional focusing of the ions is found to be accompanied by significant losses of axial kinetic energy. A Monte Carlo simulation of the energy loss process is reported that can provide agreement with the observed losses for reasonable collision cross-sections. The results suggest that operation of rf quadrupoles at relatively high pressure may find practical application in sampling ions from high (e.g., atmospheric) pressure ion sources.

PubMed Disclaimer

References

    1. Rapid Commun Mass Spectrom. 1990 Mar;4(3):81-7 - PubMed
    1. J Am Soc Mass Spectrom. 1990 Feb;1(1):53-65 - PubMed
    1. Rapid Commun Mass Spectrom. 1988 Nov;2(11):249-56 - PubMed

LinkOut - more resources