Review
. 2013 Sep;57(5):472-80.
doi: 10.4103/0019-5049.120143.

Safety features in anaesthesia machine

Affiliations
Review

Safety features in anaesthesia machine

M Subrahmanyam et al. Indian J Anaesth. 2013 Sep.

Abstract

Anaesthesia is one of the few sub-specialties of medicine, which has quickly adapted technology to improve patient safety. This application of technology can be seen in patient monitoring, advances in anaesthesia machines, intubating devices, ultrasound for visualisation of nerves and vessels, etc., Anaesthesia machines have come a long way in the last 100 years, the improvements being driven both by patient safety as well as functionality and economy of use. Incorporation of safety features in anaesthesia machines and ensuring that a proper check of the machine is done before use on a patient ensures patient safety. This review will trace all the present safety features in the machine and their evolution.

Keywords: Anaesthesia machine; hypoxia prevention; safety features.

PubMed Disclaimer

Conflict of interest statement

Conflict of Interest: None declared

Figures

Figure 1
Figure 1
The high, intermediate and low-pressure systems of the anaesthesia machine
Figure 2
Figure 2
The pin index system and its components
Figure 3
Figure 3
Safety valves
Figure 4
Figure 4
Different connectors and their safety mechanisms
Figure 5
Figure 5
Non-interchangeable screw thread
Figure 6
Figure 6
Oxygen failure warning devices
Figure 7
Figure 7
Potential unsafe and safe arrangements of flow meter tubes
Figure 8
Figure 8
Mechanical linkage of oxygen and nitrous flow meters, to prevent hypoxic mixture delivery
Figure 9
Figure 9
Transport setting in a vapourizer
Figure 10
Figure 10
Keyed and colour coded vaporiser filling systems

References

    1. Boyle HE. Improved gas-oxygen-ether outfit. Lancet. 1921;197:390.
    1. Ream AK. New directions: The anesthesia machine and the practice of anesthesia. Anesthesiology. 1978;49:307–8. - PubMed
    1. Abraham ZA, Basagoitia J. A potentially lethal anesthesia machine failure. Anesthesiology. 1987;66:589–90. - PubMed
    1. Childres WF. Malfunction of Ohio modulus anesthesia machine. Anesthesiology. 1982;56:330. - PubMed
    1. West Conshohocken, PA: ASTM International; 2005. [Last accessed on 2013 Sep 3]. ASTM Standard F1850.00, 2005, “Standard Specification for Particular Requirements for Anesthesia Workstations and Their Components,”. DOI: 10.1520/F1850-00R05. Available from: http://www.astm.org .