Biomechanical characterization of unicortical button fixation: a novel technique for proximal subpectoral biceps tenodesis
- PMID: 24253375
- DOI: 10.1007/s00167-013-2775-6
Biomechanical characterization of unicortical button fixation: a novel technique for proximal subpectoral biceps tenodesis
Abstract
Purpose: Proximal biceps tenodesis is one method for treating biceps-related pain. Tenodesis protects the length-tension relationship of the biceps muscle, maintains strength, and provides a better cosmetic appearance than tenotomy. The purpose of this investigation was to compare the mechanical properties of a unicortical metal button and an interference screw in proximal biceps tenodesis.
Methods: Six pairs of fresh-frozen shoulders were dissected, leaving the proximal biceps tendon as a free graft. On each pair of shoulders, a biceps tenodesis was performed using an interference screw or a unicortical metal button. The specimens were mounted and a cyclic load (10-60 N) was applied at 1 Hz for 200 cycles, followed by an axial load to failure. The displacement, ultimate load to failure, and mode of failure were recorded.
Results: Displacement in response to cyclic loading was 3.7 ± 2.2 mm for the interference screw and 1.9 ± 1.0 mm for the cortical button (P = 0.03). Load at failure for the interference screw was 191 ± 64 N (stiffness: 24 ± 11 N/mm) and 183 ± 61 N (stiffness: 24 ± 7. N/mm) for the unicortical button (P = n.s. for both cases).
Conclusions: As a novel technique for subpectoral biceps tenodesis, a unicortical button demonstrated significantly less displacement in response to cyclic loading than the interference screw. The ultimate load to failure and stiffness for the two methods were not different. In this way, a unicortical button may provide a reliable alternative method of fixation with a potentially lower risk of post-operative humeral fracture and a construct that permits early mobilization following biceps tenodesis.
Similar articles
-
Biomechanical Comparison of Subpectoral Biceps Tenodesis Onlay Techniques.Orthop J Sports Med. 2019 Oct 15;7(10):2325967119876276. doi: 10.1177/2325967119876276. eCollection 2019 Oct. Orthop J Sports Med. 2019. PMID: 31663008 Free PMC article.
-
Biomechanical comparison of intramedullary cortical button fixation and interference screw technique for subpectoral biceps tenodesis.Arthroscopy. 2013 May;29(5):845-53. doi: 10.1016/j.arthro.2013.01.010. Epub 2013 Mar 7. Arthroscopy. 2013. PMID: 23587927
-
Biomechanical evaluation of a unicortical button versus interference screw for subpectoral biceps tenodesis.Arthroscopy. 2013 Apr;29(4):638-44. doi: 10.1016/j.arthro.2012.11.018. Epub 2013 Feb 6. Arthroscopy. 2013. PMID: 23395466
-
A Systematic Review and Meta-analysis of Biceps Tenodesis Fixation Strengths: Fixation Type and Location Are Biomechanically Equivalent.Arthroscopy. 2020 Dec;36(12):3081-3091. doi: 10.1016/j.arthro.2020.05.055. Epub 2020 Jun 30. Arthroscopy. 2020. PMID: 32619605
-
Are Implant Choice and Surgical Approach Associated With Biceps Tenodesis Construct Strength? A Systematic Review and Meta-regression.Am J Sports Med. 2020 Apr;48(5):1273-1280. doi: 10.1177/0363546519876107. Epub 2019 Oct 4. Am J Sports Med. 2020. PMID: 31585053
Cited by
-
Clinical Outcomes of an All-Arthroscopic Biceps Tenodesis Using the Anterolateral Anchor During Concomitant Double-Row Rotator Cuff Repair.Orthop J Sports Med. 2020 Oct 9;8(10):2325967120959142. doi: 10.1177/2325967120959142. eCollection 2020 Oct. Orthop J Sports Med. 2020. PMID: 33102609 Free PMC article.
-
Management of proximal biceps tendon pathology.World J Orthop. 2022 Jan 18;13(1):36-57. doi: 10.5312/wjo.v13.i1.36. eCollection 2022 Jan 18. World J Orthop. 2022. PMID: 35096535 Free PMC article. Review.
-
[Subpectoral tenodesis of the long head of the biceps tendon with an interference screw and a cortical button : Clinical and cosmetic results after 1 year].Orthopade. 2017 Oct;46(10):864-871. doi: 10.1007/s00132-017-3458-7. Orthopade. 2017. PMID: 28808728 German.
-
Double on-lay fixation using all suture-type anchor for subpectoral biceps tenodesis has favorable functional outcomes and leads to less cosmetic deformities than single on-lay fixation.Knee Surg Sports Traumatol Arthrosc. 2019 Dec;27(12):4005-4013. doi: 10.1007/s00167-019-05663-4. Epub 2019 Aug 13. Knee Surg Sports Traumatol Arthrosc. 2019. PMID: 31410527
-
The high resistance loop (H-loop) technique used for all-inside arthroscopic knotless suprapectoral biceps tenodesis: A case series.Front Surg. 2022 Sep 14;9:917853. doi: 10.3389/fsurg.2022.917853. eCollection 2022. Front Surg. 2022. PMID: 36189379 Free PMC article.
References
-
- Basmajian JV, Latif A (1957) Integrated actions and functions of the chief flexors of the elbow: a detailed electromyographic analysis. J Bone Joint Surg Am 39-A(5):1106–1118 - PubMed
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials
Miscellaneous