Measurement of angle Kappa with Orbscan II and Galilei G4: effect of accommodation
- PMID: 24253498
- DOI: 10.1007/s00417-013-2509-y
Measurement of angle Kappa with Orbscan II and Galilei G4: effect of accommodation
Abstract
Purpose: To measure angle kappa distance with Orbscan II and Galilei G4, and to evaluate possible variations in this value for different accommodation degrees.
Methods: Angle kappa was measured using the Orbscan II and the Galilei G4 in the right eye of 80 patients aged from 20 to 40 years. This value was measured three times per eye and per device, and the average was retained. Angle kappa was measured for far vision using the Orbscan II and the Galilei G4 in a random order for each subject. The Galilei G4 was used to measure angle kappa as a function of accommodation, where the first measure started at +1 D and the vergence was changed until reaching -4 D, in 1 D steps. In both measures, the kappa distance was expressed in millimetres.
Results: At distance, the values of angle kappa were 0.43 ± 0.13 mm and 0.27 ± 0.15 mm measured with the Orbscan II and Galilei G4 systems respectively. Statistical significant differences were found (P < 0.01). With regard to the angle kappa values obtained as a function of accommodation, the values were 0.25 ± 0.15 mm, 0.26 ± 0.15 mm, 0.30 ± 0.20 mm, 0.27 ± 0.15 and 0.26 ± 0.15 mm, for +1 D, -1 D, -2 D, -3 D and -4 D respectively. No statistical significant differences were found among 0 D and the other vergences evaluated (P > 0.01).
Conclusions: For far vision, Orbscan II measured significantly higher angle kappa values than Galilei G4, the mean difference being 0.16 ± 0.08 mm. For different accommodation levels, the kappa distance did not change significantly.
Similar articles
-
Measurements of anterior chamber depth, white-to-white distance, anterior chamber angle, and pupil diameter using two Scheimpflug imaging devices.Arq Bras Oftalmol. 2014 Aug;77(4):233-7. doi: 10.5935/0004-2749.20140060. Arq Bras Oftalmol. 2014. PMID: 25410175
-
Measurement of angle kappa with synoptophore and Orbscan II in a normal population.J Refract Surg. 2007 May;23(5):456-60. doi: 10.3928/1081-597X-20070501-06. J Refract Surg. 2007. PMID: 17523505
-
Comparison of horizontal corneal diameter measurements using Galilei, EyeSys and Orbscan II systems.Clin Exp Optom. 2009 Sep;92(5):429-33. doi: 10.1111/j.1444-0938.2009.00407.x. Epub 2009 Jul 21. Clin Exp Optom. 2009. PMID: 19681922
-
Changes in anterior chamber eye during accommodation as assessed using a Dual Scheimpflug system.Arq Bras Oftalmol. 2014 Aug;77(4):243-9. doi: 10.5935/0004-2749.20140062. Arq Bras Oftalmol. 2014. PMID: 25410177
-
Comparison and reproducibility of corneal thickness and curvature readings obtained by the Galilei and the Orbscan II analysis systems.J Cataract Refract Surg. 2008 Oct;34(10):1742-7. doi: 10.1016/j.jcrs.2008.06.024. J Cataract Refract Surg. 2008. PMID: 18812127
Cited by
-
Assessment of Pupil Size and Angle Kappa in Refractive Surgery: A Population-Based Epidemiological Study in Predominantly American Caucasians.Cureus. 2023 Aug 23;15(8):e43998. doi: 10.7759/cureus.43998. eCollection 2023 Aug. Cureus. 2023. PMID: 37638275 Free PMC article.
-
Measurement of the distance between corneal apex and pupil center in patients following small-incision lenticule extraction or implantable collamer lens implantation and its correlation with the surgical-induced astigmatism.BMC Ophthalmol. 2024 Mar 7;24(1):110. doi: 10.1186/s12886-024-03352-6. BMC Ophthalmol. 2024. PMID: 38454381 Free PMC article.
-
Changes of angle Kappa and corneal morphology changes in myopic patients after Sub-Bowman-Keratomileusis.Zhong Nan Da Xue Xue Bao Yi Xue Ban. 2021 Feb 28;46(2):162-168. doi: 10.11817/j.issn.1672-7347.2021.190472. Zhong Nan Da Xue Xue Bao Yi Xue Ban. 2021. PMID: 33678653 Free PMC article. Chinese, English.
-
Image registration reveals central lens thickness minimally increases during accommodation.Clin Ophthalmol. 2017 Sep 11;11:1625-1636. doi: 10.2147/OPTH.S144238. eCollection 2017. Clin Ophthalmol. 2017. PMID: 28979092 Free PMC article.
-
Repeatability and Agreement of Chord Mu between Scheimpflug Tomography and Swept-Source Optical Coherence Tomography.Korean J Ophthalmol. 2023 Dec;37(6):510-517. doi: 10.3341/kjo.2023.0092. Epub 2023 Oct 25. Korean J Ophthalmol. 2023. PMID: 37899279 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources