Effects of flattening filter-free and volumetric-modulated arc therapy delivery on treatment efficiency
- PMID: 24257275
- PMCID: PMC5714642
- DOI: 10.1120/jacmp.v14i6.4328
Effects of flattening filter-free and volumetric-modulated arc therapy delivery on treatment efficiency
Abstract
Flattening filter-free (FFF) beams are available on an increasing number of commercial linear accelerators. FFF beams have higher dose rates than flattened beams of equivalent energy which can lead to increased efficiency of treatment delivery, especially in conjunction with increased FFF beam energy and arc-based delivery configurations. The purpose of this study is to quantify and assess the implications of improved treatment efficiency for several FFF delivery options on common types of linac applicable radiotherapy. Eleven characteristic cases representative of a variety of clinical treatment sites and prescription doses were selected from our patient population. Treatment plans were generated for a Varian TrueBeam linear accelerator. For each case, a reference plan was created using DMLC IMRT with 6MV flat beams. From the same initial objectives, plans were generated using DMLC IMRT and volumetric-modulated arc therapy (VMAT) with 6 MV FFF and 10 MV FFF beams (max. dose rates of 1400 and 2400 MU/min, respectively). The plans were delivered to a phantom; beam-on time, total treatment delivery time, monitor units (MUs), and integral dose were recorded. For plans with low dose fractionations (1.8-2.0 & 3.85 Gy/fraction), mean beam-on time difference between reference plan and most efficient FFF plan was 0.56 min (41.09% decrease); mean treatment delivery time difference between the reference plan and most efficient FFF plan was 1.54 min (range: 0.31-3.56 min), a relative improvement of 46.1% (range: 29.2%-59.2%). For plans with high dose fractionations (16-20 Gy/fraction), mean beam-on time difference was 6.79 min (74.9% decrease); mean treatment delivery time difference was 8.99 min (range: 5.40-13.05 min), a relative improvement of 71.1% (range: 53.4%- 82.4%). 10 MV FFF VMAT beams generated the most efficient plan, except in the spine SBRT case. The distribution of monitor unit counts did not vary by plan type. In cases where respiratory motion management would be applicable, 10 MV FFF DMLC IMRT reduced beam-on time/field to less than 12 sec. FFF beams significantly reduced treatment delivery time. For radiosurgical doses, the efficiency improvement for FFF beams was clinically significant. For conventional fractionation, a large improvement in relative treatment delivery time was observed, but the absolute time savings were not likely to be of clinical value. In cases that benefit from respiratory motion management, beam-on/field was reduced to a time for which most patients can comfortably maintain deep inspiratory breath hold.
Figures



Similar articles
-
MO-A-BRB-02: Facts and Fiction of Flattening Filter Free (FF-FFF) X-Rays Beams.Med Phys. 2012 Jun;39(6Part20):3861-3862. doi: 10.1118/1.4735765. Med Phys. 2012. PMID: 28517518
-
Flattening filter free in intensity-modulated radiotherapy (IMRT) - Theoretical modeling with delivery efficiency analysis.Med Phys. 2019 Jan;46(1):34-44. doi: 10.1002/mp.13267. Epub 2018 Nov 20. Med Phys. 2019. PMID: 30371944
-
Stereotactic body radiation therapy for prostate cancer: a dosimetric comparison of IMRT and VMAT using flattening filter and flattening filter-free beams.Radiat Environ Biophys. 2024 Aug;63(3):423-431. doi: 10.1007/s00411-024-01078-z. Epub 2024 Jul 6. Radiat Environ Biophys. 2024. PMID: 38969869
-
Volumetric-modulated arc therapy with RapidArc(®): An evaluation of treatment delivery efficiency.Rep Pract Oncol Radiother. 2013 Aug 17;18(6):383-6. doi: 10.1016/j.rpor.2013.07.005. Rep Pract Oncol Radiother. 2013. PMID: 24416583 Free PMC article. Review.
-
Advances in intensity-modulated radiotherapy delivery.Cancer Treat Res. 2008;139:193-214. Cancer Treat Res. 2008. PMID: 18236718 Review.
Cited by
-
Dosimetry and Feasibility Studies of Volumetric Modulated Arc Therapy With Deep Inspiration Breath-Hold Using Optical Surface Management System for Left-Sided Breast Cancer Patients.Front Oncol. 2020 Sep 3;10:1711. doi: 10.3389/fonc.2020.01711. eCollection 2020. Front Oncol. 2020. PMID: 33014848 Free PMC article.
-
Flattening Filter-Free Volumetric-Modulated Arc Radiotherapy for Left-Sided Whole-Breast, Partial-Breast, and Postmastectomy Irradiations.J Med Phys. 2022 Apr-Jun;47(2):166-172. doi: 10.4103/jmp.jmp_146_21. Epub 2022 Aug 5. J Med Phys. 2022. PMID: 36212208 Free PMC article.
-
A treatment planning study comparing Elekta VMAT and fixed field IMRT using the varian treatment planning system eclipse.Radiat Oncol. 2014 Jul 10;9:153. doi: 10.1186/1748-717X-9-153. Radiat Oncol. 2014. PMID: 25011529 Free PMC article.
-
Evaluating the use of diagnostic CT with flattening filter free beams for palliative radiotherapy: Dosimetric impact of scanner calibration variability.J Appl Clin Med Phys. 2025 Jun;26(6):e70040. doi: 10.1002/acm2.70040. Epub 2025 Feb 19. J Appl Clin Med Phys. 2025. PMID: 39969029 Free PMC article.
-
Flattening filter-free technique in volumetric modulated arc therapy for lung stereotactic body radiotherapy: A clinical comparison with the flattening filter technique.Oncol Lett. 2018 Mar;15(3):3928-3936. doi: 10.3892/ol.2018.7809. Epub 2018 Jan 16. Oncol Lett. 2018. PMID: 29563993 Free PMC article.
References
-
- Hansen HH, Connor WG, Doppke K, Boone MML. A new field flattening filter for the Clinac‐4. Radiology. 1972;103(2):443–46. - PubMed
-
- Boge RJ, Tolbert DD, Edland RW. Accessory beam flattening filter for the Varian Clinac‐4 linear accelerator. Radiology. 1975;115(2):475–77. - PubMed
-
- Fu W, Dai J, Hu Y, Han D, Song Y. Delivery time comparison for intensity‐modulated radiation therapy with/without flattening filter: a planning study. Phys Med Biol. 2004;49(8):1535–47. - PubMed
-
- Titt U, Vassiliev ON, Pönisch F, Dong L, Liu H, Mohan R. A flattening filter free photon treatment concept evaluation with Monte Carlo. Med Phys. 2006;33(6):1595–602. - PubMed
-
- Ponisch F, Titt U, Vassiliev ON, Kry SF, Mohan R. Properties of unflattened photon beams shaped by a multileaf collimator. Med Phys. 2006;33(6):1738–46. - PubMed
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials