Guidelines for anti-scatter grid use in pediatric digital radiography
Abstract
Background: Pediatric radiography presents unique challenges in balancing image quality and patient dose. Removing the anti-scatter grid reduces patient dose but also reduces image contrast. The benefit of using an anti-scatter grid decreases with decreasing patient size.
Objective: To determine patient thickness thresholds for anti-scatter grid use by comparing scatter-to-primary ratio for progressively thinner patients without a grid to the scatter-to-primary ratio for a standard adult patient with a grid.
Materials and methods: We used Solid Water™ phantoms ranging in thickness from 7 cm to 16 cm to simulate pediatric abdomens. The scatter-to-primary ratio without a grid was measured for each thickness at 60 kVp, 70 kVp and 80 kVp for X-ray fields of view (FOV) of 378 cm(2), 690 cm(2) and 1,175 cm(2) using indirect digital radiography (iDR) and computed radiography (CR). We determined thresholds for anti-scatter grid use by comparing the intersection of a fit of scatter-to-primary ratio versus patient thickness with a standard adult scatter-to-primary ratio measured for a 23-cm phantom thickness at 80 kVp with an anti-scatter grid. Dose area product (DAP) was also calculated.
Results: The scatter-to-primary ratio depended strongly on FOV and weakly on kVp; however DAP increased with decreasing kVp. Threshold thicknesses for grid use varied from 5 cm for a 14 × 17-cm FOV using iDR to 12 cm for an 8 × 10-cm FOV using computed radiography.
Conclusions: Removing the anti-scatter grid for small patients reduces patient dose without a substantial increase in scatter-to-primary ratio when the FOV is restricted appropriately. Radiologic technologists should base anti-scatter grid use on patient thickness and FOV rather than age.
Similar articles
-
Experimental evaluation of fiber-interspaced antiscatter grids for large patient imaging with digital x-ray systems.Phys Med Biol. 2007 Aug 21;52(16):4863-80. doi: 10.1088/0031-9155/52/16/010. Epub 2007 Jul 30. Phys Med Biol. 2007. PMID: 17671340
-
Evaluating the use of anti-scatter grids in adult knee radiography.Radiography (Lond). 2022 Aug;28(3):663-667. doi: 10.1016/j.radi.2022.05.004. Epub 2022 May 24. Radiography (Lond). 2022. PMID: 35623269
-
The impact of X-ray scatter correction software on abdomen radiography in terms of image quality and radiation dose.Radiography (Lond). 2024 Jul;30(4):1125-1135. doi: 10.1016/j.radi.2024.05.006. Epub 2024 May 25. Radiography (Lond). 2024. PMID: 38797045
-
The principles and effectiveness of X-ray scatter correction software for diagnostic X-ray imaging: A scoping review.Eur J Radiol. 2023 Jan;158:110600. doi: 10.1016/j.ejrad.2022.110600. Epub 2022 Nov 17. Eur J Radiol. 2023. PMID: 36444818
-
Contrast and scatter in x-ray imaging.Radiographics. 1991 Mar;11(2):307-23. doi: 10.1148/radiographics.11.2.2028065. Radiographics. 1991. PMID: 2028065 Review.
Cited by
-
Radiation dose considerations in digital radiography with an anti-scatter grid: A study using adult and pediatric phantoms.J Appl Clin Med Phys. 2023 Sep;24(9):e14081. doi: 10.1002/acm2.14081. Epub 2023 Jul 25. J Appl Clin Med Phys. 2023. PMID: 37491809 Free PMC article.
-
Evaluation of the post-processing algorithms SimGrid and S-Enhance for paediatric intensive care patients and neonates.Pediatr Radiol. 2022 May;52(6):1029-1037. doi: 10.1007/s00247-021-05279-2. Epub 2022 Feb 22. Pediatr Radiol. 2022. PMID: 35192022 Free PMC article.
-
Investigating the use of an antiscatter grid in chest radiography for average adults with a computed radiography imaging system.Br J Radiol. 2015 Mar;88(1047):20140613. doi: 10.1259/bjr.20140613. Epub 2015 Jan 9. Br J Radiol. 2015. PMID: 25571914 Free PMC article.
-
Workload and use factor data for a modern digital radiography system.J Appl Clin Med Phys. 2023 May;24(5):e13962. doi: 10.1002/acm2.13962. Epub 2023 Mar 21. J Appl Clin Med Phys. 2023. PMID: 36942718 Free PMC article.
-
Radiation use in diagnostic imaging in children: approaching the value of the pediatric radiology community.Pediatr Radiol. 2021 Apr;51(4):532-543. doi: 10.1007/s00247-020-04924-6. Epub 2021 Mar 20. Pediatr Radiol. 2021. PMID: 33743037 Review.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical