Multidetector computerized tomography enema versus magnetic resonance enema in the diagnosis of rectosigmoid endometriosis
- PMID: 24287396
- DOI: 10.1016/j.ejrad.2013.10.010
Multidetector computerized tomography enema versus magnetic resonance enema in the diagnosis of rectosigmoid endometriosis
Abstract
Purpose: To compare the accuracy of multidetector computerized tomography enema (MDCT-e) and magnetic resonance enema (MRI-e) in determining the presence of sigmoid and rectal endometriotic nodules.
Materials and methods: 260 women (32.6 ± 4.3 years) with symptoms suggestive of rectosigmoid endometriosis underwent MDCT-e and MRI-e prior to laparoscopy. After retrograde colonic distention and injection of intravenous contrast medium, patients were scanned on a 64-row MDCT scanner. MRI-e was performed on a 1.5T magnet using an 8 channels phased array coil; intestinal distention was achieved by introducing in the rectum 250-300 ml of ultrasonographic gel diluted with saline solution. Radiological findings were compared with surgical and histological results.
Results: 176 women had rectosigmoid endometriosis at surgery. There was no significant difference in the accuracy of MDCT-e (98.5%) and MRI-e (96.9%) in the diagnosis of sigmoid and rectal endometriosis (p=0.248). The sensitivity, specificity, positive predictive value, negative predictive value, positive likelihood ratio and negative likelihood ratio of MDCT-e and MRI-e were respectively 98.3%, 98.8%, 99.4%, 96.5%, 81.59, 0.02 and 97.2%, 96.4%, 98.3%, 94.1%, 26.89, 0.03.
Conclusions: Both MDCT-e and MRI-e are accurate in the diagnosis of rectal and sigmoid endometriosis.
Keywords: Bowel endometriosis; Enema; Magnetic resonance imaging; Multidetector computerized tomography; Rectum; Sigmoid.
Copyright © 2013 Elsevier Ireland Ltd. All rights reserved.