. 2014 Feb;21(1):158-65.
doi: 10.1007/s12350-013-9827-7. Epub 2013 Nov 28.

Diagnostic performance of low-dose rest/stress Tc-99m tetrofosmin myocardial perfusion SPECT using the 530c CZT camera: quantitative vs visual analysis

Affiliations

Diagnostic performance of low-dose rest/stress Tc-99m tetrofosmin myocardial perfusion SPECT using the 530c CZT camera: quantitative vs visual analysis

Fabio P Esteves et al. J Nucl Cardiol. 2014 Feb.

Abstract

Background: We set out to develop normal databases and prospectively validate abnormality criteria for a low-dose Tc-99m tetrofosmin myocardial perfusion SPECT protocol using the 530c CZT camera.

Methods: All patients received 6 mCi rest/20 mCi stress doses of Tc-99m tetrofosmin. Rest and stress images were obtained over 7-9 and 5-7 minutes according to the chest size. Low-dose CT of the chest was obtained on a standalone CT scanner. Forty patients with very low likelihood (LLK) of coronary artery disease (CAD) were used to define the normal count distributions. The abnormality criteria were prospectively validated in 55 patients who had coronary angiography and in 40 patients with LLK of CAD.

Results: The results for quantitative non-attenuation-corrected (AC) and AC analysis and visual analysis were as follows: sensitivity of 79%, 85%, and 92% (P = NS) and specificity of 44%, 75%, and 56% (P = NS), respectively. The normalcy rates for quantitative non-AC and AC analyses and visual analysis were 95%, 98%, and 98% (P = NS).

Conclusions: We have developed non-AC and AC normal databases for low-dose rest/stress Tc-99m tetrofosmin myocardial perfusion SPECT protocol using the 530c CZT camera. The per-patient diagnostic performance of quantitative analyses is not significantly different from visual analysis by an experienced reader.

PubMed Disclaimer

References

    1. Eur J Nucl Med Mol Imaging. 2013 Jul;40(7):1084-94 - PubMed
    1. J Nucl Cardiol. 2013 Oct;20(5):763-73 - PubMed
    1. J Nucl Med. 2006 Jan;47(1):51-8 - PubMed
    1. J Nucl Cardiol. 2010 Aug;17(4):709-18 - PubMed
    1. J Nucl Cardiol. 2012 Feb;19(1):19-27 - PubMed