Clinical Trial
. 2014 Apr;24(4):913-20.
doi: 10.1007/s00330-013-3075-8. Epub 2013 Dec 5.

Clinical application of bilateral high temporal and spatial resolution dynamic contrast-enhanced magnetic resonance imaging of the breast at 7 T

Affiliations
Clinical Trial

Clinical application of bilateral high temporal and spatial resolution dynamic contrast-enhanced magnetic resonance imaging of the breast at 7 T

K Pinker et al. Eur Radiol. 2014 Apr.

Abstract

Objective: The objective of our study was to evaluate the clinical application of bilateral high spatial and temporal resolution dynamic contrast-enhanced magnetic resonance imaging (HR DCE-MRI) of the breast at 7 T.

Methods: Following institutional review board approval 23 patients with a breast lesion (BIRADS 0, 4-5) were included in our prospective study. All patients underwent bilateral HR DCE-MRI of the breast at 7 T (spatial resolution of 0.7 mm(3) voxel size, temporal resolution of 14 s). Two experienced readers (r1, r2) and one less experienced reader (r3) independently assessed lesions according to BI-RADS®. Image quality, lesion conspicuity and artefacts were graded from 1 to 5. Sensitivity, specificity and diagnostic accuracy were assessed using histopathology as the standard of reference.

Results: HR DCE-MRI at 7 T revealed 29 lesions in 23 patients (sensitivity 100 % (19/19); specificity of 90 % (9/10)) resulting in a diagnostic accuracy of 96.6 % (28/29) with an AUC of 0.95. Overall image quality was excellent in the majority of cases (27/29) and examinations were not hampered by artefacts. There was excellent inter-reader agreement for diagnosis and image quality parameters (κ = 0.89-1).

Conclusion: Bilateral HR DCE-MRI of the breast at 7 T is feasible with excellent image quality in clinical practice and allows accurate breast cancer diagnosis.

Key points: • Dynamic contrast-enhanced 7-T MRI is being developed in several centres. • Bilateral high resolution DCE-MRI of the breast at 7 T is clinically applicable. • 7-T HR DCE-MRI of the breast provides excellent image quality. • 7-T HR DCE-MRI should detect breast cancer with high diagnostic accuracy.

PubMed Disclaimer