The feasibility study and characterization of a two-dimensional diode array in "magic phantom" for high dose rate brachytherapy quality assurance
- PMID: 24320410
- DOI: 10.1118/1.4822736
The feasibility study and characterization of a two-dimensional diode array in "magic phantom" for high dose rate brachytherapy quality assurance
Abstract
Purpose: High dose rate (HDR) brachytherapy is a radiation treatment technique capable of delivering large dose rates to the tumor. Radiation is delivered using remote afterloaders to drive highly active sources (commonly (192)Ir with an air KERMA strength range between 20,000 and 40,000 U, where 1 U = 1 μGy m(2)/h in air) through applicators directly into the patient's prescribed region of treatment. Due to the obvious ramifications of incorrect treatment while using such an active source, it is essential that there are methods for quality assurance (QA) that can directly and accurately verify the treatment plan and the functionality of the remote afterloader. This paper describes the feasibility study of a QA system for HDR brachytherapy using a phantom based two-dimensional 11 × 11 epitaxial diode array, named "magic phantom."
Methods: The HDR brachytherapy treatment plan is translated to the phantom with two rows of 10 (20 in total) HDR source flexible catheters, arranged above and below the diode array "magic plate" (MP). Four-dimensional source tracking in each catheter is based upon a developed fast iterative algorithm, utilizing the response of the diodes in close proximity to the (192)Ir source, sampled at 100 ms intervals by a fast data acquisition (DAQ) system. Using a (192)Ir source in a solid water phantom, the angular response of the developed epitaxial diodes utilized in the MP and also the variation of the MP response as a function of the source-to-detector distance (SDD) were investigated. These response data are then used by an iterative algorithm for source dwelling position determination. A measurement of the average transit speed between dwell positions was performed using the diodes and a fast DAQ.
Results: The angular response of the epitaxial diode showed a variation of 15% within 360°, with two flat regions above and below the detector face with less than 5% variation. For SDD distances of between 5 and 30 mm the relative response of the epitaxial diodes used in the MP is in good agreement (within 8%) with radial dose function measurements found within the TG-43 protocol, with SDD of up to 70 mm showing a 40% over response. A method for four-dimensional localization of the HDR source was developed, allowing the source dwell position to be derived within 0.50 mm of the expected position. An estimation of the average transit speed for varying step sizes was determined and was found to increase from (12.8 ± 0.3) up to (38.6 ± 0.4) cm/s for a step size of 2.5 and 50 mm, respectively.
Conclusions: Our characterization of the designed QA "magic phantom" with MP in realistic HDR photon fields demonstrates the promising performance for real-time source position tracking in four dimensions and measurements of transit times. Further development of this system will allow a full suite for QA in HDR brachytherapy and analysis, and for future in vivo tracking.
Similar articles
-
The evaluation of a 2D diode array in "magic phantom" for use in high dose rate brachytherapy pretreatment quality assurance.Med Phys. 2015 Feb;42(2):663-673. doi: 10.1118/1.4905233. Med Phys. 2015. PMID: 28102606
-
The evaluation of a 2D diode array in “magic phantom” for use in high dose rate brachytherapy pretreatment quality assurance.Med Phys. 2015 Feb;42(2):663-73. doi: 10.1118/1.4905233. Med Phys. 2015. PMID: 25771556
-
Source position verification and dosimetry in HDR brachytherapy using an EPID.Med Phys. 2013 Nov;40(11):111706. doi: 10.1118/1.4823758. Med Phys. 2013. PMID: 24320414
-
End-to-end dosimetric audit: A novel procedure developed for Irish HDR brachytherapy centres.Phys Med. 2020 Dec;80:221-229. doi: 10.1016/j.ejmp.2020.10.005. Epub 2020 Nov 13. Phys Med. 2020. PMID: 33190078 Review.
-
Determination of the dose rate around a HDR 192Ir brachytherapy source with the microDiamond and the microSilicon detector.Z Med Phys. 2023 Nov;33(4):463-478. doi: 10.1016/j.zemedi.2022.07.004. Epub 2022 Aug 26. Z Med Phys. 2023. PMID: 36038432 Free PMC article. Review.
Cited by
-
In vivo dosimetry in brachytherapy: Requirements and future directions for research, development, and clinical practice.Phys Imaging Radiat Oncol. 2020 Sep 28;16:1-11. doi: 10.1016/j.phro.2020.09.002. eCollection 2020 Oct. Phys Imaging Radiat Oncol. 2020. PMID: 33458336 Free PMC article. Review.
-
In vivo dosimetry: trends and prospects for brachytherapy.Br J Radiol. 2014 Sep;87(1041):20140206. doi: 10.1259/bjr.20140206. Epub 2014 Jul 8. Br J Radiol. 2014. PMID: 25007037 Free PMC article. Review.
-
Imaging Cherenkov emission for quality assurance of high-dose-rate brachytherapy.Sci Rep. 2020 Feb 27;10(1):3572. doi: 10.1038/s41598-020-60519-z. Sci Rep. 2020. PMID: 32108157 Free PMC article.
-
Verification of high-dose-rate brachytherapy treatment planning dose distribution using liquid-filled ionization chamber array.J Contemp Brachytherapy. 2018 Apr;10(2):142-154. doi: 10.5114/jcb.2018.75599. Epub 2018 Apr 30. J Contemp Brachytherapy. 2018. PMID: 29789763 Free PMC article.
-
Investigation of in vivo source tracking error thresholds for interstitial and intra-cavitary high-dose-rate cervical brachytherapy.J Contemp Brachytherapy. 2022 Dec;14(6):568-581. doi: 10.5114/jcb.2022.123977. Epub 2022 Dec 30. J Contemp Brachytherapy. 2022. PMID: 36819472 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous