Not all electronic foramen locators are accurate in teeth with enlarged apical foramina: an in vitro comparison of 5 brands
- PMID: 24332000
- DOI: 10.1016/j.joen.2013.09.032
Not all electronic foramen locators are accurate in teeth with enlarged apical foramina: an in vitro comparison of 5 brands
Abstract
Introduction: Electronic working length measurement during root canal treatment in teeth with enlarged apical foramina is a challenge. The aims of this in vitro study were (1) to assess the influence of foramen widening on the accuracy of 5 different electronic foramen locators (EFLs) and (2) to compare the accuracy of EFLs in different foramen sizes.
Methods: The following EFLs were used: MiniApex, Root ZXII, iPex, Propex II, and Elements Apex Locator. Each EFL was used in 3 groups (n = 20) of extracted teeth, with foramen diameters of 0.27 mm (G27), 0.47 mm (G47), and 0.72 mm (G72). Working length was measured according to manufacturer's instructions and compared with visual measurements (control method). Results were classified as accurate (equal or differences ≤ 0.05 mm) or inaccurate (differences > 0.5 mm).
Results: In G27, all EFLs yielded accurate findings (intragroup reliability; Fisher exact test, P < .05), compared with only MiniApex, Root ZXII, and Elements Apex Locator in G47 and G72. MiniApex, Root ZXII, and Elements Apex Locator were similarly accurate regardless of foramen size. iPex and Propex II were the least accurate among the devices tested, and foramen diameter influenced their accuracy, with greater diameters yielding poorer EFL performance.
Conclusions: Foramen diameter did not influence the accuracy of MiniApex, Root ZXII, and Elements Apex Locator EFLs. iPex and Propex II showed decreased accuracy as foramen size increased.
Keywords: Apical foramen; apical widening; electronic apex locator; file size; root canal instrumentation; root length determination.
Copyright © 2014 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.
Similar articles
-
In vivo accuracy of three electronic root canal length measurement devices: Dentaport ZX, Raypex 5 and ProPex II.Int Endod J. 2012 Jun;45(6):552-6. doi: 10.1111/j.1365-2591.2011.02010.x. Epub 2012 Jan 19. Int Endod J. 2012. PMID: 22257341 Clinical Trial.
-
Critical diameter of apical foramen and of file size using the Root ZX apex locator: an in vitro study.J Endod. 2011 Sep;37(9):1306-9. doi: 10.1016/j.joen.2011.05.007. Epub 2011 Jul 16. J Endod. 2011. PMID: 21846555
-
Effect of chloroform, orange solvent and eucalyptol on the accuracy of four electronic apex locators.Aust Endod J. 2013 Dec;39(3):112-5. doi: 10.1111/j.1747-4477.2011.00323.x. Epub 2011 Jul 10. Aust Endod J. 2013. PMID: 24279656
-
Determination of working length for teeth with wide or immature apices: a review.Int Endod J. 2013 Jun;46(6):483-91. doi: 10.1111/iej.12032. Epub 2012 Dec 7. Int Endod J. 2013. PMID: 23216284 Review.
-
Electronic apex locators.Int Endod J. 2004 Jul;37(7):425-37. doi: 10.1111/j.1365-2591.2004.00835.x. Int Endod J. 2004. PMID: 15189431 Review.
Cited by
-
In vitro evaluation of Root ZX and Raypex 6 in teeth with different apical diameters.J Conserv Dent. 2015 Jan-Feb;18(1):66-9. doi: 10.4103/0972-0707.148899. J Conserv Dent. 2015. PMID: 25657531 Free PMC article.
-
Consistency of electronic measurements of endodontic working length when using multiple devices from the same manufacturer-an in vitro study.Clin Oral Investig. 2018 Dec;22(9):3107-3112. doi: 10.1007/s00784-018-2404-3. Epub 2018 Mar 7. Clin Oral Investig. 2018. PMID: 29516188
-
Accuracy of Working Length Measurement Using Cone Beam Computed Tomography at Three Field of View Settings, Conventional Radiography, and Electronic Apex Locator: An Ex-vivo Study.Eur Endod J. 2024 Aug 22;9(3):266-272. doi: 10.14744/eej.2023.97769. Eur Endod J. 2024. PMID: 39102662 Free PMC article.
-
Influence of heat treatment of nickel-titanium instruments on the accuracy of an electronic apex locator integrated with endodontic motor.J Conserv Dent. 2022 Nov-Dec;25(6):596-599. doi: 10.4103/jcd.jcd_260_22. Epub 2022 Aug 17. J Conserv Dent. 2022. PMID: 36591588 Free PMC article.
-
Influence of Different Apical Foramen Morphologies on the Accuracy of Four Electronic Foramen Locators.Eur J Dent. 2024 Oct;18(4):1101-1106. doi: 10.1055/s-0044-1782214. Epub 2024 May 2. Eur J Dent. 2024. PMID: 38698613 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous