Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2014 Jan;40(1):109-12.
doi: 10.1016/j.joen.2013.09.032. Epub 2013 Oct 25.

Not all electronic foramen locators are accurate in teeth with enlarged apical foramina: an in vitro comparison of 5 brands

Affiliations
Comparative Study

Not all electronic foramen locators are accurate in teeth with enlarged apical foramina: an in vitro comparison of 5 brands

Eduardo Akisue et al. J Endod. 2014 Jan.

Abstract

Introduction: Electronic working length measurement during root canal treatment in teeth with enlarged apical foramina is a challenge. The aims of this in vitro study were (1) to assess the influence of foramen widening on the accuracy of 5 different electronic foramen locators (EFLs) and (2) to compare the accuracy of EFLs in different foramen sizes.

Methods: The following EFLs were used: MiniApex, Root ZXII, iPex, Propex II, and Elements Apex Locator. Each EFL was used in 3 groups (n = 20) of extracted teeth, with foramen diameters of 0.27 mm (G27), 0.47 mm (G47), and 0.72 mm (G72). Working length was measured according to manufacturer's instructions and compared with visual measurements (control method). Results were classified as accurate (equal or differences ≤ 0.05 mm) or inaccurate (differences > 0.5 mm).

Results: In G27, all EFLs yielded accurate findings (intragroup reliability; Fisher exact test, P < .05), compared with only MiniApex, Root ZXII, and Elements Apex Locator in G47 and G72. MiniApex, Root ZXII, and Elements Apex Locator were similarly accurate regardless of foramen size. iPex and Propex II were the least accurate among the devices tested, and foramen diameter influenced their accuracy, with greater diameters yielding poorer EFL performance.

Conclusions: Foramen diameter did not influence the accuracy of MiniApex, Root ZXII, and Elements Apex Locator EFLs. iPex and Propex II showed decreased accuracy as foramen size increased.

Keywords: Apical foramen; apical widening; electronic apex locator; file size; root canal instrumentation; root length determination.

PubMed Disclaimer

Similar articles

Cited by

Publication types

MeSH terms

LinkOut - more resources