Impact of a new respiratory amplitude-based gating technique in evaluation of upper abdominal PET lesions
- PMID: 24332351
- DOI: 10.1016/j.ejrad.2013.11.010
Impact of a new respiratory amplitude-based gating technique in evaluation of upper abdominal PET lesions
Abstract
PET acquisition requires several minutes which can lead to respiratory motion blurring, to increase partial volume effect and SUV under-estimation. To avoid these artifacts, conventional 10-min phase-based respiratory gating (PBRG) can be performed but is time-consuming and difficult with a non-compliant patient. We evaluated an automatic amplitude-based gating method (AABG) which keeps 35% of the counts at the end of expiration to minimize respiratory motion. We estimated the impact of AABG on upper abdominal lesion detectability, quantification and patient management.
Methods: We consecutively included 31 patients (82 hepatic and 25 perihepatic known lesions). Each patient underwent 3 acquisitions on a Siemens Biograph mCT (4 rings and time-of-flight): a standard free-breathing whole-body (SWB, 5-7 steps/2.5 min per step, 3.3±0.4 MBq/kg of 18F-FDG), a 10-min PBRG with six bins and a 5-min AABG method. All gated acquisitions were performed with an ANZAI respiratory gating system. SUVmax and target to background ratio (TBR, defined as the maximum SUV of the lesion divided by the mean SUV of a region of interest drawn in healthy liver) were compared.
Results: All 94 lesions in SWB images were detected in the gated images. 10-min PBRG and 5-min AABG acquisitions respectively revealed 9 and 13 new lesions and relocated 7 and 8 lesions. Four lesions revealed by 5-min AABG were missed by 10-min PBRG in 3 non-compliant patients. Both gated methods failed to relocate 2 lesions seen on SWB acquisition. Compared to SWB, TBR increased significantly with 10-min PBRG and with 5-min AABG (respectively 41±59%, p=4.10-3 and 66±75%, p=6.10-5) whereas SUVmax did not (respectively 14±43%, p=0.29 with 10-min PBRG, and 24±46%, p=0.11 with 5-min AABG).
Conclusion: The AABG is a fast and a user-friendly respiratory gating method to increase detectability and quantification of upper abdominal lesions compared to the conventional PBRG procedure and the SWB acquisition.
Keywords: PET; Respiratory gating; Respiratory motion; SUV; Upper abdominal lesions.
Copyright © 2013 Elsevier Ireland Ltd. All rights reserved.
Comment in
-
Reply to the comments on "impact of a new respiratory amplitude-based gating technique in evaluation of upper abdominal PET lesions".Eur J Radiol. 2014 Aug;83(8):1484. doi: 10.1016/j.ejrad.2014.05.024. Epub 2014 May 24. Eur J Radiol. 2014. PMID: 24917222 No abstract available.
-
Comment on "impact of a new respiratory amplitude-based gating technique in evaluation of upper abdominal PET lesions".Eur J Radiol. 2014 Aug;83(8):1483. doi: 10.1016/j.ejrad.2014.05.023. Epub 2014 May 27. Eur J Radiol. 2014. PMID: 24938670 No abstract available.
Similar articles
-
The Impact of Optimal Respiratory Gating and Image Noise on Evaluation of Intratumor Heterogeneity on 18F-FDG PET Imaging of Lung Cancer.J Nucl Med. 2016 Nov;57(11):1692-1698. doi: 10.2967/jnumed.116.173112. Epub 2016 Jun 9. J Nucl Med. 2016. PMID: 27283931
-
Comment on "impact of a new respiratory amplitude-based gating technique in evaluation of upper abdominal PET lesions".Eur J Radiol. 2014 Aug;83(8):1483. doi: 10.1016/j.ejrad.2014.05.023. Epub 2014 May 27. Eur J Radiol. 2014. PMID: 24938670 No abstract available.
-
Reply to the comments on "impact of a new respiratory amplitude-based gating technique in evaluation of upper abdominal PET lesions".Eur J Radiol. 2014 Aug;83(8):1484. doi: 10.1016/j.ejrad.2014.05.024. Epub 2014 May 24. Eur J Radiol. 2014. PMID: 24917222 No abstract available.
-
Respiratory Motion Management in PET/CT: Applications and Clinical Usefulness.Curr Radiopharm. 2017;10(2):85-92. doi: 10.2174/1874471010666170519165918. Curr Radiopharm. 2017. PMID: 28530533 Review.
-
Respiratory Gating and the Performance of PET/CT in Pulmonary Lesions.Curr Radiopharm. 2020;13(3):218-227. doi: 10.2174/1874471013666200317144629. Curr Radiopharm. 2020. PMID: 32183685 Free PMC article. Review.
Cited by
-
Simulation Study of a Frame-Based Motion Correction Algorithm for Positron Emission Imaging.Sensors (Basel). 2021 Apr 8;21(8):2608. doi: 10.3390/s21082608. Sensors (Basel). 2021. PMID: 33917742 Free PMC article.
-
Feasibility of Systematic Respiratory-Gated Acquisition in Unselected Patients Referred for 18F-Fluorodeoxyglucose Positron Emission Tomography/Computed Tomography.Front Med (Lausanne). 2018 Feb 19;5:36. doi: 10.3389/fmed.2018.00036. eCollection 2018. Front Med (Lausanne). 2018. PMID: 29516001 Free PMC article.
-
Clinical respiratory motion correction software (reconstruct, register and averaged-RRA), for 18F-FDG-PET-CT: phantom validation, practical implications and patient evaluation.Br J Radiol. 2017 Feb;90(1070):20160549. doi: 10.1259/bjr.20160549. Epub 2017 Jan 3. Br J Radiol. 2017. PMID: 27936893 Free PMC article.
-
Usefulness of Respiratory-Gated 18F-FDG PET/CT in Detecting Upper Abdominal Fever Focus.Nucl Med Mol Imaging. 2018 Oct;52(5):380-383. doi: 10.1007/s13139-018-0534-9. Epub 2018 Jul 18. Nucl Med Mol Imaging. 2018. PMID: 30344787 Free PMC article.
-
MDPET: A Unified Motion Correction and Denoising Adversarial Network for Low-Dose Gated PET.IEEE Trans Med Imaging. 2021 Nov;40(11):3154-3164. doi: 10.1109/TMI.2021.3076191. Epub 2021 Oct 27. IEEE Trans Med Imaging. 2021. PMID: 33909561 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources