Low concentration contrast medium for dual-source computed tomography coronary angiography by a combination of iterative reconstruction and low-tube-voltage technique: feasibility study
- PMID: 24332352
- DOI: 10.1016/j.ejrad.2013.11.006
Low concentration contrast medium for dual-source computed tomography coronary angiography by a combination of iterative reconstruction and low-tube-voltage technique: feasibility study
Abstract
Objectives: To assess the impact of low-concentration contrast medium on vascular enhancement, image quality and radiation dose of coronary CT angiography (cCTA) by using a combination of iterative reconstruction (IR) and low-tube-voltage technique.
Materials and methods: One hundred patients were prospectively randomized to two types of contrast medium and underwent prospective electrocardiogram-triggering cCTA (Definition Flash, Siemens Healthcare; collimation: 128 mm × 0.6mm; tube current: 300 mAs). Fifty patients received Iopromide 370 were scanned using the conventional tube setting (100 kVp or 120 kVp if BMI ≥ 25 kg/m(2)) and reconstructed with filtered back projection (FBP). Fifty patients received Iodixanol 270 were scanned using the low-tube-voltage (80 kVp or 100 kVp if BMI ≥ 25 kg/m(2)) technique and reconstructed with IR. CT attenuation was measured in coronary artery and other anatomical regions. Noise, image quality and radiation dose were compared.
Results: Compared with two Iopromide 370 subgroups, Iomeprol 270 subgroups showed no significant difference in CT attenuation (576.63 ± 95.50 vs. 569.51 ± 118.93 for BMI< 25 kg/m(2), p=0.647 and 394.19 ± 68.09 vs. 383.72 ± 63.11 for BMI ≥ 25 kg/m(2), p=0.212), noise (in various anatomical regions of interest) and image quality (3.5 vs. 4.0, p=0.13), but significantly (0.41 ± 0.17 vs. 0.94 ± 0.45 for BMI< 25 kg/m(2), p<0.001 and 1.14 ± 0.24 vs. 2.37 ± 0.69 for BMI ≥ 25 kg/m(2), p<0.001) lower radiation dose, which reflects dose saving of 56.4% and 51.9%, respectively.
Conclusions: Combined IR with low-tube-voltage technique, a low-concentration contrast medium of 270 mg I/ml can still maintain the contrast enhancement without impairing image quality, as well as significantly lower the radiation dose.
Keywords: Contrast medium concentration; Coronary CT angiography; Image quality; Iterative reconstruction; Radiation dose.
Copyright © 2013 Elsevier Ireland Ltd. All rights reserved.
Similar articles
-
Low-concentration contrast medium for 128-slice dual-source CT coronary angiography at a very low radiation dose using prospectively ECG-triggered high-pitch spiral acquisition.Acad Radiol. 2015 Feb;22(2):195-202. doi: 10.1016/j.acra.2014.07.025. Epub 2014 Nov 11. Acad Radiol. 2015. PMID: 25457735 Clinical Trial.
-
Effect of reduced x-ray tube voltage, low iodine concentration contrast medium, and sinogram-affirmed iterative reconstruction on image quality and radiation dose at coronary CT angiography: results of the prospective multicenter REALISE trial.J Cardiovasc Comput Tomogr. 2015 May-Jun;9(3):215-24. doi: 10.1016/j.jcct.2015.01.010. Epub 2015 Jan 22. J Cardiovasc Comput Tomogr. 2015. PMID: 25843243 Clinical Trial.
-
Low contrast and radiation dose coronary CT angiography using a 320-row system and a refined contrast injection and timing method.J Cardiovasc Comput Tomogr. 2015 Jan-Feb;9(1):19-27. doi: 10.1016/j.jcct.2014.12.002. Epub 2014 Dec 24. J Cardiovasc Comput Tomogr. 2015. PMID: 25677790 Clinical Trial.
-
Current and Novel Imaging Techniques in Coronary CT.Radiographics. 2015 Jul-Aug;35(4):991-1010. doi: 10.1148/rg.2015140181. Epub 2015 Jun 5. Radiographics. 2015. PMID: 26046942 Review.
-
New horizons in cardiac CT.Clin Radiol. 2016 Aug;71(8):758-67. doi: 10.1016/j.crad.2016.01.022. Epub 2016 Feb 28. Clin Radiol. 2016. PMID: 26932775 Review.
Cited by
-
Personalized application of three different concentrations of iodinated contrast media in coronary computed tomography angiography.J Cell Mol Med. 2020 May;24(10):5446-5453. doi: 10.1111/jcmm.15196. Epub 2020 Mar 30. J Cell Mol Med. 2020. PMID: 32227625 Free PMC article.
-
Dual Energy CT Angiography of Peripheral Arterial Disease: Feasibility of Using Lower Contrast Medium Volume.PLoS One. 2015 Sep 29;10(9):e0139275. doi: 10.1371/journal.pone.0139275. eCollection 2015. PLoS One. 2015. PMID: 26418007 Free PMC article. Clinical Trial.
-
Use of coronary artery calcium scanning beyond coronary computed tomographic angiography in the emergency department evaluation for acute chest pain: the ROMICAT II trial.Circ Cardiovasc Imaging. 2015 Mar;8(3):e002225. doi: 10.1161/CIRCIMAGING.114.002225. Epub 2015 Feb 20. Circ Cardiovasc Imaging. 2015. PMID: 25710925 Free PMC article. Clinical Trial.
-
Guideline on the use of iodinated contrast media in patients with kidney disease 2018.Jpn J Radiol. 2020 Jan;38(1):3-46. doi: 10.1007/s11604-019-00850-2. Jpn J Radiol. 2020. PMID: 31709498 No abstract available.
-
Feasibility of dual-low scheme combined with iterative reconstruction technique in acute cerebral infarction volume CT whole brain perfusion imaging.Exp Ther Med. 2017 Jul;14(1):163-168. doi: 10.3892/etm.2017.4451. Epub 2017 May 11. Exp Ther Med. 2017. PMID: 28672909 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous