Comparative Study
. 2013 Dec;58 Suppl 1(0 1):e119-22.
doi: 10.1016/j.jcv.2013.09.018.

Relative accuracy of serum, whole blood, and oral fluid HIV tests among Seattle men who have sex with men

Affiliations
Comparative Study

Relative accuracy of serum, whole blood, and oral fluid HIV tests among Seattle men who have sex with men

Joanne D Stekler et al. J Clin Virol. 2013 Dec.

Abstract

Background: Point-of-care (POC) rapid HIV tests have sensitivity during the "window period" comparable only to earliest generation EIAs. To date, it is unclear whether any POC test performs significantly better than others.

Objective: Compare abilities of POC tests to detect early infection in real time.

Study design: Men who have sex with men (MSM) were recruited into a prospective, cross-sectional study at two HIV testing sites and a research clinic. Procedures compared four POC tests: one performed on oral fluids and three on fingerstick whole blood specimens. Specimens from participants with negative POC results were tested by EIA and pooled nucleic acid amplification testing (NAAT). McNemar's exact tests compared numbers of HIV-infected participants detected.

Results: Between February 2010 and May 2013, 104 men tested HIV-positive during 2479 visits. Eighty-two participants had concordant reactive POC results, 3 participants had concordant non-reactive POC tests but reactive EIAs, and 8 participants had acute infection. Of 12 participants with discordant POC results, OraQuick ADVANCE Rapid HIV-1/2 Antibody Test performed on oral fluids identified fewer infections than OraQuick performed on fingerstick (p = .005), Uni-Gold Recombigen HIV test (p = .01), and determine HIV-1/2 Ag/Ab combo (p = .005).

Conclusions: These data confirm that oral fluid POC testing detects fewer infections than other methods and is best reserved for circumstances precluding fingerstick or venipuncture. Regardless of specimen type, POC tests failed to identify many HIV-infected MSM in Seattle. In populations with high HIV incidence, the currently approved POC antibody tests are inadequate unless supplemented with p24 antigen tests or NAAT.

Keywords: HIV testing; Oral fluids; Rapid HIV test.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Stekler J, Swenson PD, Wood RW, Handsfield HH, Golden MR. Targeted screening for primary HIV infection through pooled HIV-RNA testing in men who have sex with men. AIDS. 2005;19:1323–1325. - PubMed
    1. Fiebig EW, Wright DJ, Rawal BD, Garrett PE, Schumacher RT, Peddada L, et al. Dynamics of HIV viremia and antibody seroconversion in plasma donors: implications for diagnosis and staging of primary HIV infection. Aids. 2003;17:1871–1879. - PubMed
    1. Wawer MJ, Gray RH, Sewankambo NK, Serwadda D, Li X, Laeyendecker O, et al. Rates of HIV-1 transmission per coital act, by stage of HIV-1 infection, in Rakai, Uganda. Journal of Infectious Diseases. 2005;191:1403–1409. - PubMed
    1. Stekler JD, Swenson PD, Coombs RW, Dragavon J, Thomas KK, Brennan CA, et al. HIV testing in a high-incidence population: is antibody testing alone good enough? Clinical Infectious Diseases. 2009;49:444–453. - PMC - PubMed
    1. Stekler J, Wood RW, Swenson PD, Golden M. Negative rapid HIV antibody testing during early HIV infection. Annals of Internal Medicine. 2007;147:147–148. - PubMed

Publication types