Safety, efficacy, and predictability of laser in situ keratomileusis to correct myopia or myopic astigmatism with a 750 Hz scanning-spot laser system

Affiliations

Safety, efficacy, and predictability of laser in situ keratomileusis to correct myopia or myopic astigmatism with a 750 Hz scanning-spot laser system

Minoru Tomita et al. J Cataract Refract Surg. 2014 Feb.

Abstract

Purpose: To evaluate the clinical outcomes of laser in situ keratomileusis (LASIK) to correct myopia or myopic astigmatism using the Amaris 750S 750 Hz excimer laser.

Setting: Private LASIK center, Tokyo, Japan.

Design: Case series.

Methods: Patients with myopia or myopic astigmatism (spherical equivalent -0.50 to -11.63 diopters [D]), a corrected distance visual acuity (CDVA) of 20/20 or better, and an estimated residual bed thickness of 300 μm or more had LASIK using the aspheric aberration-free ablation profile of the 750 Hz scanning-spot laser and the Femto LDV Crystal Line femtosecond laser for flap creation. Study parameters included uncorrected distance visual acuity (UDVA), CDVA, manifest refraction, astigmatism, and higher-order aberrations (HOAs).

Results: The study included 1280 eyes (685 patients). At 3 months, 96.6% of eyes had a UDVA of 20/20 or better and 99.1% had 20/32 or better; 94.1% of eyes were within ± 0.50 D of the intended correction and 98.9% were within ± 1.00 D; 89.7% of eyes had no residual cylinder and 96.0% had a postoperative astigmatism of less than 0.50 D. All eyes had a postoperative CDVA of 20/20 or better. The HOAs increased postoperatively (P<.001), with mean total postoperative corneal and ocular HOAs of 0.66 μm ± 0.20 (SD) and 0.56 ± 0.23 μm, respectively. The efficacy index and safety index were 1.02 and 1.06, respectively.

Conclusion: Laser in situ keratomileusis with the 750 Hz scanning-spot laser was safe, effective, and predictable. No specific clinical side effects that might be associated with a high repetition rate occurred.

Financial disclosure: Mr. Magnago is an employee of Schwind eye-tech-solutions GmbH. No other author has a financial or proprietary interest in any material or method mentioned.