An in vitro comparison of root canal length determination by DentaPort ZX and iPex apex locators
- PMID: 24347893
- PMCID: PMC3842727
- DOI: 10.4103/0972-0707.120953
An in vitro comparison of root canal length determination by DentaPort ZX and iPex apex locators
Abstract
Aim: THE AIM OF THE PRESENT STUDY IS TO EVALUATE AND TO COMPARE THE ACCURACY OF TWO ELECTRONIC APEX LOCATORS (EALS): DentaPort ZX and iPex, at a position 0.5 mm short of the apical foramen.
Materials and methods: Thirty single-rooted, mandibular premolar teeth were selected. Standard access cavities were prepared and the actual length (AL) was calculated. The samples were then embedded in alginate and the electronic measurements were determined and recorded.
Results: The results obtained showed that in determining the root canal length with a tolerance level of ±0.5 mm, i.e., AL ± 0.5 mm, DentaPort ZX was accurate in 93.3% of the samples and iPex was accurate in 90% of the samples at a position 0.5 mm short of the apical foramen.
Conclusion: A strong correlation was seen between the two electronic methods and AL and also in between the two EALs, showing the possibility of their use to measure the root canal length. No statistically significant difference was found between both the apex locators.
Keywords: Actual length; DentaPort ZX; apex locator; iPex.
Conflict of interest statement
Similar articles
-
Evaluation of the accuracy of different apex locators in determining the working length during root canal retreatment.J Dent Res Dent Clin Dent Prospects. 2020 Spring;14(2):125-129. doi: 10.34172/joddd.2020.026. Epub 2020 Jun 17. J Dent Res Dent Clin Dent Prospects. 2020. PMID: 32908654 Free PMC article.
-
The Effect of Long Service Life on the Accuracy of Dentaport ZX Electronic Apex Locator.Iran Endod J. 2022;17(4):195-199. doi: 10.22037/iej.v17i4.34623. Iran Endod J. 2022. PMID: 36703688 Free PMC article.
-
In vivo evaluation of the iPex and Root ZX electronic apex locators using various irrigants.Int Endod J. 2013 Aug;46(8):769-74. doi: 10.1111/iej.12057. Epub 2013 Mar 28. Int Endod J. 2013. PMID: 23551276 Clinical Trial.
-
Comparative Evaluation of Accuracy of Ipex, Root Zx Mini, and Epex Pro Apex Locators in Teeth with Artificially Created Root Perforations in Presence of Various Intracanal Irrigants.Eur Endod J. 2020 Mar 19;5(1):6-9. doi: 10.14744/eej.2019.07279. eCollection 2020. Eur Endod J. 2020. PMID: 32342031 Free PMC article.
-
Evaluation of the accuracy of electronic apex locators Dentaport ZX and Ray Pex 4 under clinical conditions.Folia Med (Plovdiv). 2007;49(3-4):75-9. Folia Med (Plovdiv). 2007. PMID: 18504939
Cited by
-
Influence of root canal curvature on the accuracy of an electronic apex locator: An in vitro study.J Conserv Dent. 2014 Nov;17(6):583-6. doi: 10.4103/0972-0707.144610. J Conserv Dent. 2014. PMID: 25506150 Free PMC article.
-
Comparison of Accuracy of Dentaport ZX, Rootor and E-Pex Pro Electronic Apex Locators in Two Simulated Clinical Conditions: An In Vitro Study.J Conserv Dent. 2022 Jan-Feb;25(1):58-62. doi: 10.4103/jcd.jcd_557_21. Epub 2022 May 2. J Conserv Dent. 2022. PMID: 35722068 Free PMC article.
-
A comparative assessment of the accuracies of Raypex 5, Raypex 6, iPex and iPex II electronic apex locators: An in vitro study.J Istanb Univ Fac Dent. 2017 Jan 2;51(1):28-33. doi: 10.17096/jiufd.61309. eCollection 2017. J Istanb Univ Fac Dent. 2017. PMID: 28955583 Free PMC article.
-
Evaluation of the accuracy of different apex locators in determining the working length during root canal retreatment.J Dent Res Dent Clin Dent Prospects. 2020 Spring;14(2):125-129. doi: 10.34172/joddd.2020.026. Epub 2020 Jun 17. J Dent Res Dent Clin Dent Prospects. 2020. PMID: 32908654 Free PMC article.
-
Accuracy of two electronic apex locators in locating root perforations in curved canals in dry and wet conditions: A comparative in vitro study.Dent Res J (Isfahan). 2019 Nov 12;16(6):407-412. eCollection 2019 Nov-Dec. Dent Res J (Isfahan). 2019. PMID: 31803387 Free PMC article.
References
-
- Ricucci D, Langeland K. Apical limit of root canal instrumentation and obturation, part 2. A histological study. Int Endod J. 1998;31:394–409. - PubMed
-
- Dummer PM, McGinn JH, Rees DG. The position and topography of the apical canal constriction and apical foramen. Int Endod J. 1984;17:192–8. - PubMed
-
- Sjogren U, Hagglund B, Sundqvist G, Wing K. Factors affecting the long-term results of endodontic treatment. J Endod. 1990;16:498–504. - PubMed
-
- Suzuki K. Experimental study on iontophoresis. J Jap Stomatol. 1942;16:411–29.
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials