Evaluation of a Novel Cone Beam Computed Tomography Scanner for Bone Density Examinations in Preoperative 3D Reconstructions and Correlation with Primary Implant Stability
- PMID: 24373386
- DOI: 10.1111/cid.12193
Evaluation of a Novel Cone Beam Computed Tomography Scanner for Bone Density Examinations in Preoperative 3D Reconstructions and Correlation with Primary Implant Stability
Abstract
Background: Bone density examination (BDE) using preoperative cone beam computed tomography (CBCT) might be used to predict primary implant stability in implant patients.
Purpose: The aim of the study was to validate a novel CBCT scanner in vitro with regard to BDE in preoperative scans and to analyze the in vivo correlation of CBCT scan results with primary implant stability measurements.
Materials and methods: A CBCT scanner was validated in vitro with regard to spatial uniformity and linearity of CT numbers (Hounsfield units, HU) by using a series of phantoms and plastic and hydroxyapatite specimens of various densities. Forty-nine patients (27 female and 22 male, mean age 55.6 ± 9.8) were scanned prior to and 1 to 6 months after the placement of 155 dental implants of different lengths and diameters. Mean and peak insertion torque (IT) in Ncm were registered during implant placement. Resonance frequency analysis (RFA) measurements in ISQ units were performed after placement. The second scan was used to export and superimpose the exact positions of bone and implants into the first scan. Virtual probes with the same length as the actual implant were automatically placed at the implant sites, and mean HU values were measured in a 1 mm-wide circular corridor from the tip of the threads and out.
Results: The in vitro validation showed high uniformity and linearity of CT numbers (HU). The clinical study showed significant correlations between bone density and ISQ, mean IT, and peak IT, respectively.
Conclusions: The WhiteFox CBCT scanner measures bone density with high accuracy. There is a correlation between bone density and primary implant stability as assessed with IT and RFA measurements. The findings suggest that BDE may be used as an additional feature in treatment-planning software to estimate primary stability at predetermined implant sites.
Keywords: CBCT; bone density; clinical study; implant stability; insertion torque; resonance frequency.
© 2013 Wiley Periodicals, Inc.
Similar articles
-
Relationships between bone density values from cone beam computed tomography, maximum insertion torque, and resonance frequency analysis at implant placement: a pilot study.Int J Oral Maxillofac Implants. 2011 Sep-Oct;26(5):1051-6. Int J Oral Maxillofac Implants. 2011. PMID: 22010089
-
Short communication: use of a diagnostic software to predict bone density and implant stability in preoperative CTs.Clin Implant Dent Relat Res. 2012 Aug;14(4):553-7. doi: 10.1111/j.1708-8208.2010.00291.x. Epub 2010 Jun 25. Clin Implant Dent Relat Res. 2012. PMID: 20586783
-
Correlation of cone beam CT-derived bone density parameters with primary implant stability assessed by peak insertion torque and periotest in the maxilla.J Craniomaxillofac Surg. 2019 Mar;47(3):461-467. doi: 10.1016/j.jcms.2019.01.002. Epub 2019 Jan 10. J Craniomaxillofac Surg. 2019. PMID: 30683621
-
Accuracy of linear measurements on CBCT images related to presurgical implant treatment planning: A systematic review.Clin Oral Implants Res. 2018 Oct;29 Suppl 16:393-415. doi: 10.1111/clr.13142. Clin Oral Implants Res. 2018. PMID: 30328204
-
Cone beam computed tomography in implant dentistry: recommendations for clinical use.BMC Oral Health. 2018 May 15;18(1):88. doi: 10.1186/s12903-018-0523-5. BMC Oral Health. 2018. PMID: 29764458 Free PMC article. Review.
Cited by
-
Stress Distribution Analysis of Threaded Implants for Digital Dentistry.Int J Environ Res Public Health. 2022 Oct 4;19(19):12674. doi: 10.3390/ijerph191912674. Int J Environ Res Public Health. 2022. PMID: 36231974 Free PMC article.
-
Influence of time on primary stability of dental implants placed with osteotomes due to the elastic properties of peri-implant bone.J Dent Sci. 2019 Dec;14(4):358-364. doi: 10.1016/j.jds.2019.04.004. Epub 2019 May 21. J Dent Sci. 2019. PMID: 31890122 Free PMC article.
-
Buccal bone thickness at dental implants in the maxillary anterior region with large bony defects at time of immediate implant placement: A 1-year cohort study.Clin Implant Dent Relat Res. 2019 Feb;21(1):73-79. doi: 10.1111/cid.12701. Epub 2018 Dec 11. Clin Implant Dent Relat Res. 2019. PMID: 30548160 Free PMC article.
-
Factors Influencing Resonance Frequency Analysis (RFA) Measurements and 5-Year Survival of Neoss Dental Implants.Int J Dent. 2019 Apr 1;2019:3209872. doi: 10.1155/2019/3209872. eCollection 2019. Int J Dent. 2019. PMID: 31065267 Free PMC article.
-
Inter- and intraobserver reproducibility of buccal bone measurements at dental implants with cone beam computed tomography in the esthetic region.Int J Implant Dent. 2015 Dec;1(1):8. doi: 10.1186/s40729-015-0007-1. Epub 2015 Apr 18. Int J Implant Dent. 2015. PMID: 27747630 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical