Diagnostic performance of optical coherence tomography ganglion cell--inner plexiform layer thickness measurements in early glaucoma

Affiliations

Diagnostic performance of optical coherence tomography ganglion cell--inner plexiform layer thickness measurements in early glaucoma

Jean-Claude Mwanza et al. Ophthalmology. 2014 Apr.

Abstract

Purpose: To evaluate the glaucoma diagnostic performance of ganglion cell inner-plexiform layer (GCIPL) parameters used individually and in combination with retinal nerve fiber layer (RNFL) or optic nerve head (ONH) parameters measured with Cirrus HD-OCT (Carl Zeiss Meditec, Inc, Dublin, CA).

Design: Prospective cross-sectional study.

Participants: Fifty patients with early perimetric glaucoma and 49 age-matched healthy subjects.

Methods: Three peripapillary RNFL and 3 macular GCIPL scans were obtained in 1 eye of each participant. A patient was considered glaucomatous if at least 2 of the 3 RNFL or GCIPL scans had the average or at least 1 sector measurement flagged at 1% to 5% or less than 1%. The diagnostic performance was determined for each GCIPL, RNFL, and ONH parameter as well as for binary or-logic and and-logic combinations of GCIPL with RNFL or ONH parameters.

Main outcome measures: Sensitivity, specificity, positive likelihood ratio (PLR), and negative likelihood ratio (NLR).

Results: Among GCIPL parameters, the minimum had the best diagnostic performance (sensitivity, 82.0%; specificity, 87.8%; PLR, 6.69; and NLR, 0.21). Inferior quadrant was the best RNFL parameter (sensitivity, 74%; specificity, 95.9%; PLR, 18.13; and NLR, 0.27), as was rim area (sensitivity, 68%; specificity, 98%; PLR, 33.3; and NLR, 0.33) among ONH parameters. The or-logic combination of minimum GCIPL and average RNFL provided the overall best diagnostic performance (sensitivity, 94%; specificity, 85.7%; PRL, 6.58; and NLR, 0.07) as compared with the best RNFL, best ONH, and best and-logic combination (minimum GCIPL and inferior quadrant RNFL; sensitivity, 64%; specificity, 100%; PLR, infinity; and NPR, 0.36).

Conclusions: The binary or-logic combination of minimum GCIPL and average RNFL or rim area provides better diagnostic performances than those of and-logic combinations or best single GCIPL, RNFL, or ONH parameters. This finding may be clinically valuable for the diagnosis of early glaucoma.