Pulmonary hemorrhage after cryoballoon ablation for pulmonary vein isolation in the treatment of atrial fibrillation
- PMID: 24394827
- DOI: 10.1378/chest.13-0761
Pulmonary hemorrhage after cryoballoon ablation for pulmonary vein isolation in the treatment of atrial fibrillation
Abstract
Pulmonary vein isolation has evolved over the past years as an alternative for the treatment of symptomatic recurrences of atrial fibrillation refractory to antiarrhythmic drug treatment. Both radiofrequency energy and cryoballoon ablation have proven useful in this setting. We present the case of a 55-year-old male patient undergoing cryoballoon ablation complicated with pulmonary hemorrhage. The cause of this rare complication may be found in the damage of vascular venous structures near the ablation zone or, alternatively, in hemorrhagic damage of the pulmonary vein surrounding tissue (or less probably to direct injury of the lingular bronchus). The extremely low temperatures achieved in this case (which are often associated with deep balloon position inside the veins) are alarming and should alert the physician about the possibility of an excessively intrapulmonary vein deployment of the cryoablation balloon.
Comment in
-
Hemoptysis: déjà vu for cryoballoon use for pulmonary vein isolation for atrial fibrillation ablation.Chest. 2014 Jun;145(6):1435. doi: 10.1378/chest.14-0236. Chest. 2014. PMID: 24889450 No abstract available.