Assessment of patient setup errors in IGRT in combination with a six degrees of freedom couch
- PMID: 24418323
- DOI: 10.1016/j.zemedi.2013.11.002
Assessment of patient setup errors in IGRT in combination with a six degrees of freedom couch
Abstract
Purpose: The range of patient setup errors in six dimensions detected in clinical routine for cranial as well as for extracranial treatments, were analyzed while performing linear accelerator based stereotactic treatments with frameless patient setup systems. Additionally, the need for re-verification of the patient setup for situations where couch rotations are involved was analyzed for patients treated in the cranial region.
Methods and materials: A total of 2185 initial (i.e. after pre-positioning the patient with the infrared system but before image guidance) patient setup errors (1705 in the cranial and 480 in the extracranial region) obtained by using ExacTrac (BrainLAB AG, Feldkirchen, Germany) were analyzed. Additionally, the patient setup errors as a function of the couch rotation angle were obtained by analyzing 242 setup errors in the cranial region. Before the couch was rotated, the patient setup error was corrected at couch rotation angle 0° with the aid of image guidance and the six degrees of freedom (6DoF) couch. For both situations attainment rates for two different tolerances (tolerance A: ± 0.5mm, ± 0.5°; tolerance B: ± 1.0 mm, ± 1.0°) were calculated.
Results: The mean (± one standard deviation) initial patient setup errors for the cranial cases were -0.24 ± 1.21°, -0.23 ± 0.91° and -0.03 ± 1.07° for the pitch, roll and couch rotation axes and 0.10 ± 1.17 mm, 0.10 ± 1.62 mm and 0.11 ± 1.29 mm for the lateral, longitudinal and vertical axes, respectively. Attainment rate (all six axes simultaneously) for tolerance A was 0.6% and 13.1% for tolerance B, respectively. For the extracranial cases the corresponding values were -0.21 ± 0.95°, -0.05 ± 1.08° and -0.14 ± 1.02° for the pitch, roll and couch rotation axes and 0.15 ± 1.77 mm, 0.62 ± 1.94 mm and -0.40 ± 2.15 mm for the lateral, longitudinal and vertical axes. Attainment rate (all six axes simultaneously) for tolerance A was 0.0% and 3.1% for tolerance B, respectively. After initial setup correction and rotation of the couch to treatment position a re-correction has to be performed in 77.4% of all cases to fulfill tolerance A and in 15.6% of all cases to fulfill tolerance B.
Conclusion: The analysis of the data shows that all six axes of a 6DoF couch are used extensively for patient setup in clinical routine. In order to fulfill high patient setup accuracies (e.g. for stereotactic treatments), a 6DoF couch is recommended. Moreover, re-verification of the patient setup after rotating the couch is required in clinical routine.
Keywords: 6D-Tisch; 6DoF couch; IGRT; Patient setup errors; Patienten-Lagerungsunsicherheit; Stereotaktische Radiotherapie; stereotactic radiotherapy.
Copyright © 2014. Published by Elsevier GmbH.
Similar articles
-
Clinical evaluation of a robotic 6-degree of freedom treatment couch for frameless radiosurgery.Int J Radiat Oncol Biol Phys. 2012 May 1;83(1):467-74. doi: 10.1016/j.ijrobp.2011.05.048. Epub 2011 Sep 22. Int J Radiat Oncol Biol Phys. 2012. PMID: 21945110
-
Evaluation of a new six degrees of freedom couch for radiation therapy.Med Phys. 2013 Nov;40(11):111710. doi: 10.1118/1.4823789. Med Phys. 2013. PMID: 24320418
-
Residual setup errors in cranial stereotactic radiosurgery without six degree of freedom robotic couch: Frameless versus rigid immobilization systems.J Appl Clin Med Phys. 2020 Mar;21(3):87-93. doi: 10.1002/acm2.12828. Epub 2020 Feb 18. J Appl Clin Med Phys. 2020. PMID: 32068342 Free PMC article.
-
IGRT and motion management during lung SBRT delivery.Phys Med. 2017 Dec;44:113-122. doi: 10.1016/j.ejmp.2017.06.006. Epub 2017 Jun 21. Phys Med. 2017. PMID: 28647449 Review.
-
Daily stereotactic ultrasound prostate targeting: inter-user variability.Technol Cancer Res Treat. 2003 Apr;2(2):161-70. doi: 10.1177/153303460300200213. Technol Cancer Res Treat. 2003. PMID: 12680798 Review.
Cited by
-
A simplified non-coplanar volumetric modulated arc therapy for the whole brain radiotherapy with hippocampus avoidance.Front Oncol. 2023 Apr 17;13:1143564. doi: 10.3389/fonc.2023.1143564. eCollection 2023. Front Oncol. 2023. PMID: 37152035 Free PMC article.
-
Comprehensive Analysis of Set-Up Gain of 6-Dimensional Cone-Beam CT Correction Method in Radiotherapy for Head and Neck and Brain Tumors.Comput Math Methods Med. 2022 Oct 20;2022:2964023. doi: 10.1155/2022/2964023. eCollection 2022. Comput Math Methods Med. 2022. PMID: 36311255 Free PMC article.
-
Mask dependency of the lacrimal gland dose under whole brain radiotherapy when the six-degrees of freedom couch is not available.J Appl Clin Med Phys. 2023 Jul;24(7):e14052. doi: 10.1002/acm2.14052. Epub 2023 May 31. J Appl Clin Med Phys. 2023. PMID: 37254709 Free PMC article.
-
Positioning accuracy during VMAT of gynecologic malignancies and the resulting dosimetric impact by a 6-degree-of-freedom couch in combination with daily kilovoltage cone beam computed tomography.Radiat Oncol. 2015 Apr 26;10:104. doi: 10.1186/s13014-015-0412-x. Radiat Oncol. 2015. PMID: 25927659 Free PMC article.
-
Commissioning and Evaluation of a Third-Party 6 Degrees-of-Freedom Couch Used in Radiotherapy.Technol Cancer Res Treat. 2019 Jan 1;18:1533033819870778. doi: 10.1177/1533033819870778. Technol Cancer Res Treat. 2019. PMID: 31434547 Free PMC article.