Xpert MTB/RIF testing in a low tuberculosis incidence, high-resource setting: limitations in accuracy and clinical impact
- PMID: 24429440
- DOI: 10.1093/cid/ciu022
Xpert MTB/RIF testing in a low tuberculosis incidence, high-resource setting: limitations in accuracy and clinical impact
Abstract
Background: Xpert MTB/RIF, the first automated molecular test for tuberculosis, is transforming the diagnostic landscape in low-income countries. However, little information is available on its performance in low-incidence, high-resource countries.
Methods: We evaluated the accuracy of Xpert in a university hospital tuberculosis clinic in Montreal, Canada, for the detection of pulmonary tuberculosis on induced sputum samples, using mycobacterial cultures as the reference standard. We also assessed the potential reduction in time to diagnosis and treatment initiation.
Results: We enrolled 502 consecutive patients who presented for evaluation of possible active tuberculosis (most with abnormal chest radiographs, only 18% symptomatic). Twenty-five subjects were identified to have active tuberculosis by culture. Xpert had a sensitivity of 46% (95% confidence interval [CI], 26%-67%) and specificity of 100% (95% CI, 99%-100%) for detection of Mycobacterium tuberculosis. Sensitivity was 86% (95% CI, 42%-100%) in the 7 subjects with smear-positive results, and 28% (95% CI, 10%-56%) in the remaining subjects with smear-negative, culture-positive results; in this latter group, positive Xpert results were obtained a median 12 days before culture results. Subjects with positive cultures but negative Xpert results had minimal disease: 11 of 13 had no symptoms on presentation, and mean time to positive liquid culture results was 28 days (95% CI, 25-47 days) compared with 14 days (95% CI, 8-21 days) in Xpert/culture-positive cases.
Conclusions: Our findings suggest limited potential impact of Xpert testing in high-resource, low-incidence ambulatory settings due to lower sensitivity in the context of less extensive disease, and limited potential to expedite diagnosis beyond what is achieved with the existing, well-performing diagnostic algorithm.
Keywords: diagnostics; molecular testing; point-of-care; tuberculosis.
Comment in
-
Molecular laboratory testing for tuberculosis: innovators, early adopters, or laggards?Clin Infect Dis. 2014 Apr;58(7):977-9. doi: 10.1093/cid/ciu024. Epub 2014 Jan 14. Clin Infect Dis. 2014. PMID: 24429435 No abstract available.
-
Performance of Xpert MTB/RIF in induced sputum samples at detecting Mycobacterium tuberculosis.Clin Infect Dis. 2014 Sep 15;59(6):911-2. doi: 10.1093/cid/ciu375. Epub 2014 May 22. Clin Infect Dis. 2014. PMID: 24855147 No abstract available.
-
Reply to El Sahly.Clin Infect Dis. 2014 Sep 15;59(6):912. doi: 10.1093/cid/ciu376. Epub 2014 May 22. Clin Infect Dis. 2014. PMID: 24855148 No abstract available.
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical