Migration characteristics and early clinical results of a novel-finned press-fit acetabular cup
- PMID: 24442859
- DOI: 10.1007/s00508-013-0488-y
Migration characteristics and early clinical results of a novel-finned press-fit acetabular cup
Abstract
Background: Ana Nova® is a novel-finned press-fit acetabular cup which showed superior biomechanical characteristics in an experimental set-up. Using Einzel Bild Röntgen Analyse (EBRA) measurements should offer the opportunity to predict implant survival at an early stage. The purpose of this study was to assess migration and clinical outcome 2 years after total hip replacement by a novel-finned press-fit acetabular cup.
Methods: In this study, migration and clinical results of the implant were prospectively assessed in 67 patients. Clinical outcome was assessed using the Harris hip score (HHS). Migration analyses were performed using the computer assisted EBRA system. Data were analyzed for normal distribution using the Kolmogorov-Smirnov test. Group comparisons were performed using the analysis of variance (ANOVA) test. P-values less than 0.05 were considered statistically significant.
Results: At 2 years after surgery, none of the implants needed revision and HHS increased from 39.7 up to 92.2. In contrast to the beneficial clinical outcome, 17 of 44 patients showed increased total migration ( 1 mm/2a).
Conclusions: Adverse migration data in this study might predict aseptic loosening and decreased survival of the implant. According to previous studies, it is possible that this effect occurred because of limited accuracy of the EBRA system. In our opinion, migration analyses may not be recommended as a screening tool in a 2 year follow-up.
Similar articles
-
Migration characteristics and early clinical results of the NANOS® short-stem hip arthroplasty.Wien Klin Wochenschr. 2015 May;127(9-10):375-8. doi: 10.1007/s00508-015-0756-0. Epub 2015 Mar 5. Wien Klin Wochenschr. 2015. PMID: 25739650
-
No Benefit After THA Performed With Computer-assisted Cup Placement: 10-year Results of a Randomized Controlled Study.Clin Orthop Relat Res. 2016 Oct;474(10):2085-93. doi: 10.1007/s11999-016-4863-7. Clin Orthop Relat Res. 2016. PMID: 27150344 Free PMC article. Clinical Trial.
-
Acetabular revision arthroplasty using press-fitted jumbo cups: an average 10-year follow-up study.Arch Orthop Trauma Surg. 2019 Aug;139(8):1149-1160. doi: 10.1007/s00402-019-03214-7. Epub 2019 Jun 11. Arch Orthop Trauma Surg. 2019. PMID: 31187257
-
Outcome of short- to medium-term migration analysis of a cementless short stem total hip arthroplasty using EBRA-FCA: a radiological and clinical study.Arch Orthop Trauma Surg. 2020 Feb;140(2):247-253. doi: 10.1007/s00402-019-03315-3. Epub 2019 Nov 30. Arch Orthop Trauma Surg. 2020. PMID: 31786645
-
Subsidence of a metaphyseal-anchored press-fit stem after 4-year follow-up: an EBRA-FCA analysis.Arch Orthop Trauma Surg. 2022 Aug;142(8):2075-2082. doi: 10.1007/s00402-021-04068-8. Epub 2021 Jul 21. Arch Orthop Trauma Surg. 2022. PMID: 34287700 Free PMC article. Review.
Cited by
-
Migration characteristics and early clinical results of the NANOS® short-stem hip arthroplasty.Wien Klin Wochenschr. 2015 May;127(9-10):375-8. doi: 10.1007/s00508-015-0756-0. Epub 2015 Mar 5. Wien Klin Wochenschr. 2015. PMID: 25739650
-
Frequency of safety signals from scientific reports, manufactures, registers, and other sources for a random selection of hip and knee prostheses.Acta Orthop. 2025 Jun 25;96:460-466. doi: 10.2340/17453674.2025.44035. Acta Orthop. 2025. PMID: 40557841 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical