Migration characteristics and early clinical results of a novel-finned press-fit acetabular cup

Affiliations
Clinical Trial

Migration characteristics and early clinical results of a novel-finned press-fit acetabular cup

Martin Kaipel et al. Wien Klin Wochenschr. 2014 Apr.

Abstract

Background: Ana Nova® is a novel-finned press-fit acetabular cup which showed superior biomechanical characteristics in an experimental set-up. Using Einzel Bild Röntgen Analyse (EBRA) measurements should offer the opportunity to predict implant survival at an early stage. The purpose of this study was to assess migration and clinical outcome 2 years after total hip replacement by a novel-finned press-fit acetabular cup.

Methods: In this study, migration and clinical results of the implant were prospectively assessed in 67 patients. Clinical outcome was assessed using the Harris hip score (HHS). Migration analyses were performed using the computer assisted EBRA system. Data were analyzed for normal distribution using the Kolmogorov-Smirnov test. Group comparisons were performed using the analysis of variance (ANOVA) test. P-values less than 0.05 were considered statistically significant.

Results: At 2 years after surgery, none of the implants needed revision and HHS increased from 39.7 up to 92.2. In contrast to the beneficial clinical outcome, 17 of 44 patients showed increased total migration ( 1 mm/2a).

Conclusions: Adverse migration data in this study might predict aseptic loosening and decreased survival of the implant. According to previous studies, it is possible that this effect occurred because of limited accuracy of the EBRA system. In our opinion, migration analyses may not be recommended as a screening tool in a 2 year follow-up.

PubMed Disclaimer

Similar articles

Cited by

References

    1. J Bone Joint Surg Br. 1996 May;78(3):422-6 - PubMed
    1. Int Orthop. 2008 Dec;32(6):791-4 - PubMed
    1. Int Orthop. 2005 Feb;29(1):14-7 - PubMed
    1. J Arthroplasty. 2004 Jun;19(4 Suppl 1):92-4 - PubMed
    1. J Biomech. 1995 Oct;28(10):1225-36 - PubMed

Publication types

MeSH terms

LinkOut - more resources