. 2014 Dec;18(9):2067-75.
doi: 10.1007/s00784-014-1186-5. Epub 2014 Jan 18.

Laboratory validation of a new gas-enhanced dentine liquid permeation evaluation system

Affiliations
Free article

Laboratory validation of a new gas-enhanced dentine liquid permeation evaluation system

Anas Al-Jadaa et al. Clin Oral Investig. 2014 Dec.
Free article

Abstract

Aim: To validate a new automated dentine permeability testing platform based on pressure change measurements.

Methodology: A split chamber was designed allowing for concomitant measurement of fluid permeation and pressure difference. In a first test, system reliability was assessed by interposing a solid metal disk, embedded composite resin disks, or teeth by consecutively measuring eight times under standardized conditions. Secondly, the repeatability and applicability of the method was tested in a dentine wound model by using intact third molars: Class I (2 × 5 mm) and a full occlusal preparation as well a ceramic restoration were consecutively performed and repeatedly measured eight times each. In the last test, the system detection limit as well correlation between gas pressure difference and liquid permeation were evaluated: Again, third molars were used and occlusal preparations of increasing size (2 × 5, 3 × 5, 4 × 5, and 5 × 5 mm and full occlusal preparations, respectively) were made. Data was analyzed for the linearity of measurement, and R (2) values were calculated.

Results: The embedding procedure allowed for perfect separation of the two chambers, and no significant variation in repeated measurements of evaluated samples for the respective treatments (p = 0.05) was found. The detection was 0.002 hPa/min for the pressure slope and 0.0225 μl/min for the fluid infiltration, respectively. The saline volume was highly correlating to the gas pressure changes (R (2) = 0.996, p < 0.0001).

Conclusions: The presented method is a reliable and exact tool to assess dentine permeability by nondestructive and repeatable measurements.

Clinical relevance: This method is suitable for measurements and comparison of the effectiveness of dentine wounds sealing materials.

PubMed Disclaimer

Similar articles

Cited by

References

    1. J Dent Res. 1974 Nov-Dec;53(6):1503 - PubMed
    1. J Biomed Mater Res. 2000 Sep 15;51(4):650-9 - PubMed
    1. Arch Oral Biol. 1996 Nov;41(11):1065-72 - PubMed
    1. J Dent Res. 1983 Sep;62(9):956-9 - PubMed
    1. Clin Oral Investig. 2013 Mar;17 Suppl 1:S9-19 - PubMed

Publication types

LinkOut - more resources