Evaluation of radiation doses delivered in different chest CT protocols
- PMID: 24454417
- PMCID: PMC3894921
- DOI: 10.12659/PJR.889952
Evaluation of radiation doses delivered in different chest CT protocols
Abstract
Background: There are differences in the reference diagnostic levels for the computed tomography (CT) of the chest as cited in different literature sources. The doses are expressed either in weighted CT dose index (CTDIVOL) used to express the dose per slice, dose-length product (DLP), and effective dose (E). The purpose of this study was to assess the radiation dose used in Low Dose Computer Tomography (LDCT) of the chest in comparison with routine chest CT examinations as well as to compare doses delivered in low dose chest CT with chest X-ray doses.
Material/methods: CTDIVOL and DLP doses were taken to analysis from routine CT chest examinations (64 MDCT TK LIGHT SPEED GE Medical System) performed in 202 adult patients with FBP reconstruction: 51 low dose, 106 helical, 20 angio CT, and 25 high resolution CT protocols, as well as 19 helical protocols with iterative ASIR reconstruction. The analysis of chest X-ray doses was made on the basis of reports from 44 examinations.
Results: Mean values of CTDIVOL and DLP were, respectively: 2.1 mGy and 85.1 mGy·cm, for low dose, 9.7 mGy and 392.3 mGy·cm for helical, 18.2 mGy and 813.9 mGy·cm for angio CT, 2.3 mGy and 64.4 mGy·cm for high resolution CT, 8.9 mGy. and 317.6 mGy·cm for helical ASIR protocols. Significantly lower CTDIVOL and DLP values were observed for low dose and high resolution CT versus the remaining CT protocols; doses delivered in CT ASIR protocols were also lower (80-81%). The ratio between medial doses in low dose CT and chest X-ray was 11.56.
Conclusions: Radiation dose in extended chest LDCT with parameters allowing for identification of mediastinal structures and adrenal glands is still much lower than that in standard CT protocols. Effective doses predicted for LDCT may exceed those used in chest X-ray examinations by a factor of 4 to 12, depending on LDCT scan parameters. Our results, as well as results from other authors, suggest a possibility of reducing the dose by means of iterative reconstruction. Efforts towards further dose reduction which would permit replacing chest X-ray with low dose CT in certain research screening projects should be encouraged.
Keywords: chest CT; computed tomography (CT); low dose CT; lung CT; radiation protection; radiation safety.
Similar articles
-
Adult Computed Tomography examinations in Uganda: Towards determining the National Diagnostic Reference Levels.BMC Med Imaging. 2022 Jun 11;22(1):112. doi: 10.1186/s12880-022-00838-x. BMC Med Imaging. 2022. PMID: 35690743 Free PMC article.
-
Establishment of a local diagnostic reference level for computed tomography chest and abdomen in two different cities in Saudi Arabia.Appl Radiat Isot. 2024 Feb;204:111147. doi: 10.1016/j.apradiso.2023.111147. Epub 2023 Dec 14. Appl Radiat Isot. 2024. PMID: 38113663
-
Radiation doses from head, neck, chest and abdominal CT examinations: an institutional dose report.Diagn Interv Radiol. 2021 Jan;27(1):147-151. doi: 10.5152/dir.2020.19560. Diagn Interv Radiol. 2021. PMID: 33475510 Free PMC article.
-
Low-Dose Chest CT Protocols for Imaging COVID-19 Pneumonia: Technique Parameters and Radiation Dose.Life (Basel). 2023 Apr 12;13(4):992. doi: 10.3390/life13040992. Life (Basel). 2023. PMID: 37109522 Free PMC article. Review.
-
A systematic review and meta-analysis on the radiation dose of computed tomography in hybrid nuclear medicine imaging.EJNMMI Phys. 2023 May 25;10(1):32. doi: 10.1186/s40658-023-00553-8. EJNMMI Phys. 2023. PMID: 37227561 Free PMC article. Review.
Cited by
-
Letting go of what we believe about radiation and the risk of cancer in children.Pediatr Radiol. 2017 Jan;47(1):113-115. doi: 10.1007/s00247-016-3697-5. Epub 2016 Oct 14. Pediatr Radiol. 2017. PMID: 27743004 No abstract available.
-
Pilot study of low-dose nonenhanced computed tomography with iterative reconstruction for diagnosis of urinary stones.Korean J Urol. 2014 Sep;55(9):581-6. doi: 10.4111/kju.2014.55.9.581. Epub 2014 Sep 5. Korean J Urol. 2014. PMID: 25237459 Free PMC article.
-
Multi-factorial considerations for intra-thoracic lymph node evaluations of healthy cats on computed tomographic images.BMC Vet Res. 2021 Jan 28;17(1):59. doi: 10.1186/s12917-021-02771-7. BMC Vet Res. 2021. PMID: 33509167 Free PMC article.
-
Assessment of Effective Dose Associated with Coronary Computed Tomography Angiography in Isfahan Province, Iran.J Med Signals Sens. 2018 Jan-Mar;8(1):60-64. J Med Signals Sens. 2018. PMID: 29535926 Free PMC article.
-
Tensor Gradient L₀-Norm Minimization-Based Low-Dose CT and Its Application to COVID-19.IEEE Trans Instrum Meas. 2021 Jan 19;70:4503012. doi: 10.1109/TIM.2021.3050190. eCollection 2021. IEEE Trans Instrum Meas. 2021. PMID: 35582003 Free PMC article.
References
-
- Tack D. Radiation Safety in Thoracic Imaging. European Cardiology. 2009;7:19–22.
-
- Tsapaki V, Kottou S, Papadimitriou D. Application of European Commission reference dose levels in CT examinations in Crete, Greece. Br J Radiol. 2001;74:836–40. - PubMed
-
- Katsura M, Matsuda I, Akahane M, et al. Model-based iterative reconstruction technique for radiation dose reduction in chest CT: comparison with the adaptive statistical iterative reconstruction technique. Eur Radiol. 2012;22:1613–23. - PubMed
-
- Kalra MK, Maher MM, Toth TL, et al. Strategies for CT radiation dose optimization. Radiology. 2004;230:619–28. - PubMed