Performance of an automated six-wavelength photometer (Radiometer OSM3) for routine measurement of hemoglobin derivatives
- PMID: 2448064
Performance of an automated six-wavelength photometer (Radiometer OSM3) for routine measurement of hemoglobin derivatives
Abstract
A new instrument for spectrophotometric determination of oxygen saturation (SO2) and the fractions of carboxyhemoglobin (FHbCO) and methemoglobin (FHi), the Hemoximeter OSM3 (Radiometer), was tested by comparative measurements with a multiwavelength method. We found that it gave reliable results for freshly drawn samples from patients. When samples with artificially induced high proportions of HbCO or Hi, or both, were introduced, results were still sufficiently accurate. The presence of sulfhemoglobin (SHb) seriously interfered with the measurement of SO2, FHbCO, and FHi, but the instrument reliably indicated its presence. Because SHb is rarely encountered, a reliable warning for its presence is adequate for clinical practice. Using the instrument in its fetal mode, we obtained accurate results for blood samples from newborns.
Similar articles
-
A new trend in blood gas chemistry: the measurement of clinically relevant hemoglobin derivatives. Performance of the OSM3 hemoximeter.Scand J Clin Lab Invest Suppl. 1987;188:57-60. Scand J Clin Lab Invest Suppl. 1987. PMID: 3482481
-
Accurate measurements of hemoglobin oxygen saturation, and fractions of carboxyhemoglobin and methemoglobin in fetal blood using Radiometer OSM3: corrections for fetal hemoglobin fraction and pH.Scand J Clin Lab Invest Suppl. 1990;203:235-9. doi: 10.3109/00365519009087516. Scand J Clin Lab Invest Suppl. 1990. PMID: 1708524
-
Multicomponent analysis of hemoglobin derivatives with reversed-optics spectrophotometer.Clin Chem. 1984 Mar;30(3):373-9. Clin Chem. 1984. PMID: 6697482
-
Spectrophotometry of hemoglobin and hemoglobin derivatives.Adv Clin Chem. 1983;23:199-257. doi: 10.1016/s0065-2423(08)60401-1. Adv Clin Chem. 1983. PMID: 6398614 Review. No abstract available.
-
Chemicals reacting with various forms of hemoglobin: biological significance, mechanisms, and determination.J Forensic Sci. 1991 May;36(3):662-72. J Forensic Sci. 1991. PMID: 1856636 Review.
Cited by
-
Hemoglobin determination with paired emitter detector diode.Anal Bioanal Chem. 2011 Mar;399(9):3293-7. doi: 10.1007/s00216-010-4358-4. Epub 2010 Nov 3. Anal Bioanal Chem. 2011. PMID: 21046074 Free PMC article.
-
Feasible and simple exclusion criteria for pulmonary reference populations.Thorax. 2007 Sep;62(9):792-8. doi: 10.1136/thx.2006.071480. Epub 2007 Mar 27. Thorax. 2007. PMID: 17389756 Free PMC article.
-
Case files of the University of Cincinnati fellowship in medical toxicology: two patients with acute lethal occupational exposure to hydrogen sulfide.J Med Toxicol. 2007 Jun;3(2):73-81. doi: 10.1007/BF03160912. J Med Toxicol. 2007. PMID: 18072164 Free PMC article. No abstract available.
Publication types
MeSH terms
Substances
LinkOut - more resources
Other Literature Sources