Evaluation of cone-beam computed tomography in the diagnosis of simulated small osseous defects in the mandibular condyle
- PMID: 24485728
- DOI: 10.1016/j.ajodo.2013.10.014
Evaluation of cone-beam computed tomography in the diagnosis of simulated small osseous defects in the mandibular condyle
Abstract
Introduction: In this study, we investigated the impact of defect size and scan voxel size on the accuracy of cone-beam computed tomography (CBCT) diagnoses of simulated condylar defects and assessed the value of orthodontic CBCT images typically scanned at lower settings (0.4-mm voxel size and full-size field of view) in diagnosing condylar erosion defects.
Methods: Cylindrical holes simulating condylar defects with varied diameters (≤2, 2-3, and >3 mm) and depths (≤2 and >2 mm) were created in 22 fresh pig mandibular condyles, with defect number and size per condyle and quadrant randomly determined. With the soft tissues repositioned, 2 CBCT scans (voxel sizes, 0.4 and 0.2 mm) of the pig heads were obtained from an i-CAT unit (Imaging Science International, Hatfield, Pa). Reconstructed CBCT data were analyzed independently by 2 calibrated, blinded raters using Dolphin-3D (Dolphin Imaging and Management Solutions, Chatsworth, Calif) for defect identification and localization and defect diameter and depth measurements, which were compared with physical diagnoses obtained from polyvinyl siloxane impressions.
Results: Identification and localization of simulated defects demonstrated moderate interrater reliability and excellent specificity and sensitivity, except for extremely small defects (both diameter and depth ≤2 mm) viewed with 0.4-mm scans, which had a significantly lower sensitivity (67.3%). Geometric measurements of simulated defects demonstrated good but not excellent interrater reliability and submillimeter inaccuracy for all defects. Receiver operating characteristic analyses demonstrated that the overall accuracy of diagnosing simulated condylar defects based on CBCT geometric measurements was fair and good for the 0.4-mm and 0.2-mm voxel-size scans, respectively. With the prevalence of condylar erosion defects in the patients considered, the positive predictive values of diagnoses based on 0.5-mm size (diameter or depth) cutoff points were near 15% and 50% for asymptomatic and symptomatic temporomandibular joints, respectively; the negative predictive values were near 95% and 90%, respectively.
Conclusions: When using orthodontic CBCT images for diagnosing condylar osseous defects, extremely small (<2 mm) defects can be difficult to detect; caution is also needed for the diagnostic accuracy of positive diagnoses, especially those from asymptomatic temporomandibular joints.
Copyright © 2014 American Association of Orthodontists. Published by Mosby, Inc. All rights reserved.
Similar articles
-
Comparison of cone-beam computed tomography with multislice computed tomography in detection of small osseous condylar defects.Am J Orthod Dentofacial Orthop. 2016 Jul;150(1):130-9. doi: 10.1016/j.ajodo.2015.12.019. Am J Orthod Dentofacial Orthop. 2016. PMID: 27364215
-
Detection accuracy of condylar defects in cone beam CT images scanned with different resolutions and units.Dentomaxillofac Radiol. 2014;43(3):20130414. doi: 10.1259/dmfr.20130414. Epub 2014 Jan 22. Dentomaxillofac Radiol. 2014. PMID: 24408818 Free PMC article.
-
Factors affecting the accuracy of buccal alveolar bone height measurements from cone-beam computed tomography images.Am J Orthod Dentofacial Orthop. 2013 Mar;143(3):353-63. doi: 10.1016/j.ajodo.2012.10.019. Am J Orthod Dentofacial Orthop. 2013. PMID: 23452969
-
Application of cone beam computed tomography for assessment of the temporomandibular joints.Aust Dent J. 2012 Mar;57 Suppl 1:109-18. doi: 10.1111/j.1834-7819.2011.01663.x. Aust Dent J. 2012. PMID: 22376103 Review.
-
Reliability and accuracy of segmentation of mandibular condyles from different three-dimensional imaging modalities: a systematic review.Dentomaxillofac Radiol. 2020 Jul;49(5):20190150. doi: 10.1259/dmfr.20190150. Epub 2019 Dec 3. Dentomaxillofac Radiol. 2020. PMID: 31778321 Free PMC article.
Cited by
-
[Development and validation of a clinical automatic diagnosis system based on diagnostic criteria for temporomandibular disorders].Beijing Da Xue Xue Bao Yi Xue Ban. 2025 Feb 18;57(1):192-201. doi: 10.19723/j.issn.1671-167X.2025.01.029. Beijing Da Xue Xue Bao Yi Xue Ban. 2025. PMID: 39856527 Free PMC article. Chinese.
-
Effect of Voxel Size and Object Location in the Field of View on Detection of Bone Defects in Cone Beam Computed Tomography.J Dent (Tehran). 2016 Aug;13(4):279-286. J Dent (Tehran). 2016. PMID: 28127320 Free PMC article.
-
Cone beam computed tomography-based models versus multislice spiral computed tomography-based models for assessing condylar morphology.Oral Surg Oral Med Oral Pathol Oral Radiol. 2016 Jan;121(1):96-105. doi: 10.1016/j.oooo.2015.10.015. Epub 2015 Oct 20. Oral Surg Oral Med Oral Pathol Oral Radiol. 2016. PMID: 26679363 Free PMC article.
-
Validation of Osteoarthritis synthetic defect database via non-rigid registration.Proc SPIE Int Soc Opt Eng. 2015 Mar 17;9417:94171C. doi: 10.1117/12.2081983. Proc SPIE Int Soc Opt Eng. 2015. PMID: 26236073 Free PMC article.
-
Diagnosis of simulated condylar bone defects using panoramic radiography, spiral tomography and cone-beam computed tomography: A comparison study.J Clin Exp Dent. 2015 Feb 1;7(1):e34-9. doi: 10.4317/jced.51736. eCollection 2015 Feb. J Clin Exp Dent. 2015. PMID: 25810839 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous