A limited-angle intrafraction verification (LIVE) system for radiation therapy
- PMID: 24506590
- DOI: 10.1118/1.4861820
A limited-angle intrafraction verification (LIVE) system for radiation therapy
Abstract
Purpose: Currently, no 3D or 4D volumetric x-ray imaging techniques are available for intrafraction verification of target position during actual treatment delivery or in-between treatment beams, which is critical for stereotactic radiosurgery (SRS) and stereotactic body radiation therapy (SBRT) treatments. This study aims to develop a limited-angle intrafraction verification (LIVE) system to use prior information, deformation models, and limited angle kV-MV projections to verify target position intrafractionally.
Methods: The LIVE system acquires limited-angle kV projections simultaneously during arc treatment delivery or in-between static 3D/IMRT treatment beams as the gantry moves from one beam to the next. Orthogonal limited-angle MV projections are acquired from the beam's eye view (BEV) exit fluence of arc treatment beam or in-between static beams to provide additional anatomical information. MV projections are converted to kV projections using a linear conversion function. Patient prior planning CT at one phase is used as the prior information, and the on-board patient volume is considered as a deformation of the prior images. The deformation field is solved using the data fidelity constraint, a breathing motion model extracted from the planning 4D-CT based on principal component analysis (PCA) and a free-form deformation (FD) model. LIVE was evaluated using a 4D digital extended cardiac torso phantom (XCAT) and a CIRS 008A dynamic thoracic phantom. In the XCAT study, patient breathing pattern and tumor size changes were simulated from CT to treatment position. In the CIRS phantom study, the artificial target in the lung region experienced both size change and position shift from CT to treatment position. Varian Truebeam research mode was used to acquire kV and MV projections simultaneously during the delivery of a dynamic conformal arc plan. The reconstruction accuracy was evaluated by calculating the 3D volume percentage difference (VPD) and the center of mass (COM) difference of the tumor in the true on-board images and reconstructed images.
Results: In both simulation and phantom studies, LIVE achieved substantially better reconstruction accuracy than reconstruction using PCA or FD deformation model alone. In the XCAT study, the average VPD and COM differences among different patient scenarios for LIVE system using orthogonal 30° scan angles were 4.3% and 0.3 mm when using kV+BEV MV. Reducing scan angle to 15° increased the average VPD and COM differences to 15.1% and 1.7 mm. In the CIRS phantom study, the VPD and COM differences for the LIVE system using orthogonal 30° scan angles were 6.4% and 1.4 mm. Reducing scan angle to 15° increased the VPD and COM differences to 51.9% and 3.8 mm.
Conclusions: The LIVE system has the potential to substantially improve intrafraction target localization accuracy by providing volumetric verification of tumor position simultaneously during arc treatment delivery or in-between static treatment beams. With this improvement, LIVE opens up a new avenue for margin reduction and dose escalation in both fractionated treatments and SRS and SBRT treatments.
Similar articles
-
Reducing scan angle using adaptive prior knowledge for a limited-angle intrafraction verification (LIVE) system for conformal arc radiotherapy.Phys Med Biol. 2017 May 7;62(9):3859-3882. doi: 10.1088/1361-6560/aa6913. Epub 2017 Mar 24. Phys Med Biol. 2017. PMID: 28338470 Free PMC article.
-
Image acquisition optimization of a limited-angle intrafraction verification (LIVE) system for lung radiotherapy.Med Phys. 2018 Jan;45(1):340-351. doi: 10.1002/mp.12647. Epub 2017 Nov 30. Med Phys. 2018. PMID: 29091287 Free PMC article.
-
A technique for estimating 4D-CBCT using prior knowledge and limited-angle projections.Med Phys. 2013 Dec;40(12):121701. doi: 10.1118/1.4825097. Med Phys. 2013. PMID: 24320487
-
Dynamic targeting image-guided radiotherapy.Med Dosim. 2006 Summer;31(2):113-25. doi: 10.1016/j.meddos.2005.12.014. Med Dosim. 2006. PMID: 16690452 Review.
-
Image guidance: treatment target localization systems.Front Radiat Ther Oncol. 2007;40:72-93. doi: 10.1159/000106029. Front Radiat Ther Oncol. 2007. PMID: 17641503 Review.
Cited by
-
Dosimetric verification of lung cancer treatment using the CBCTs estimated from limited-angle on-board projections.Med Phys. 2015 Aug;42(8):4783-95. doi: 10.1118/1.4926559. Med Phys. 2015. PMID: 26233206 Free PMC article.
-
Advanced 4-dimensional cone-beam computed tomography reconstruction by combining motion estimation, motion-compensated reconstruction, biomechanical modeling and deep learning.Vis Comput Ind Biomed Art. 2019;2(1):23. doi: 10.1186/s42492-019-0033-6. Epub 2019 Dec 12. Vis Comput Ind Biomed Art. 2019. PMID: 32190409 Free PMC article.
-
Iterative inversion of deformation vector fields with feedback control.Med Phys. 2018 Jul;45(7):3147-3160. doi: 10.1002/mp.12962. Epub 2018 Jun 10. Med Phys. 2018. PMID: 29757473 Free PMC article.
-
Daily edge deformation prediction using an unsupervised convolutional neural network model for low dose prior contour based total variation CBCT reconstruction (PCTV-CNN).Biomed Phys Eng Express. 2019 Oct;5(6):065013. doi: 10.1088/2057-1976/ab446b. Epub 2019 Oct 7. Biomed Phys Eng Express. 2019. PMID: 32587754 Free PMC article.
-
Utilizing the TrueBeam Advanced Imaging Package to monitor intrafraction motion with periodic kV imaging and automatic marker detection during VMAT prostate treatments.J Appl Clin Med Phys. 2020 Mar;21(3):184-191. doi: 10.1002/acm2.12822. Epub 2020 Jan 24. J Appl Clin Med Phys. 2020. PMID: 31981305 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources