Evaluation of automated immunoassays for 25(OH)-vitamin D determination in different critical populations before and after standardization of the assays

Affiliations

Evaluation of automated immunoassays for 25(OH)-vitamin D determination in different critical populations before and after standardization of the assays

Etienne Cavalier et al. Clin Chim Acta. .

Abstract

Introduction: Standardization of immunoassays for 25(OH)-vitamin D determination is a major problem in clinical practice. A worldwide standardization program has started to address this and will reduce the bias observed between immunoassays. We aimed to calibrate 5 immunoassays on a LC-MS/MS traceable to the SRM 2972 and the ID-LC-MS/MS 25(OH)D Reference Method Procedure to see if the re-standardization would be efficient in a population of 3rd trimester pregnant women (PW), hemodialysis (HD) and osteoporosis (OP) patient.

Material and methods: 184 serum samples (25(OH)D: 8.4-87 ng/ml) were selected to calibrate the immunoassays (Abbott-Architect, Roche-Elecsys, DiaSorin-Liaison, Siemens-Centaur and IDS-iSYS). Chromsystems MassChrom method was used as the referenced. Serum obtained in 34 PW, 25 HD and 34 OP patients were used as comparatives.

Results: After adjusting to LC-MS/MS, immunoassays had regression slopes nearly identical to 1.0 with intercepts <0.5 ng/ml. However, in special populations, a systematic bias was still observed, except for iSYS.

Conclusions: Re-standardization of 25(OH)D immunoassay will globally improve the differences. However, patients with a different serum matrix will still present significantly different results when they will be run with different methods. For those patients, the LC-MS/MS method seems to be the method of choice, even if some immunoassays are less influenced than others.

Keywords: Standardization; Uremia; Vitamin D; Vitamin D–binding protein.