. 2014 Jul;21(7):1116-22.
doi: 10.1016/j.jocn.2013.11.010. Epub 2013 Dec 7.

Delayed complications after flow-diverter stenting: reactive in-stent stenosis and creeping stents

Affiliations

Delayed complications after flow-diverter stenting: reactive in-stent stenosis and creeping stents

José E Cohen et al. J Clin Neurosci. 2014 Jul.

Abstract

We assessed the frequency and severity of changes in stent configuration and location after the treatment of intracranial aneurysms, and patterns of in-stent stenosis. We retrospectively reviewed data for consecutive aneurysm patients managed with endovascular implantation of flow-diverter stents (Silk Flow Diverter [Balt Extrusion, Montmorency, France] and Pipeline Embolization Device [ev3/Coviden, Minneapolis, MN, USA]) from October 2011 to July 2012. Routine 2, 6, 9-12, and 16-20 month follow-up angiograms were compared, with a focus on changes in stent configuration and location from immediately after deployment to angiographic follow-up, and the incidence and development of in-stent stenosis. Thirty-four patients with 42 aneurysms met inclusion criteria. The Silk device was implanted in 16 patients (47%, single device in 15), the Pipeline device in 18 (53%, single device in 16). On first follow-up angiography, in-stent stenosis was observed in 38% of Silk devices and 39% of Pipeline devices. In-stent stenosis was asymptomatic in 12 of 13 patients. One woman presented with transient ischemic attacks and required stent angioplasty due to end tapering and mild, diffuse in-stent stenosis. Configuration and location changes, including stent creeping and end tapering were seen in 2/16 patients (13%) with Silk devices, and 0/18 patients with Pipeline devices. We describe stent creeping and end tapering as unusual findings with the potential for delayed clinical complications. In-stent stenosis, with a unique behavior, is a frequent angiographic finding observed after flow-diverter stent implant. The stenosis is usually asymptomatic; however, close clinical and angiographic monitoring is mandatory for individualized management.

Keywords: Brain aneurysm; Complications; Flow-diverter stent; In-stent stenosis; Stent migration.